文献速递腹部泌尿生殖技术与安全CT

腹部盆腔单源快速千伏切换能谱CT中真实平扫与虚拟平扫衰减值的比较

Comparison of true unenhanced and virtual unenhanced (VUE) attenuation values in abdominopelvic single-source rapid kilovoltage-switching spectral CT.
PubMed 文献 · Abdom Radiol (NY) · 2017-01-01

回顾性分析94例腹盆三相CT(肝脏方案47例、胰腺方案47例),比较真实平扫与基于快速千伏切换双能CT动脉期及静脉期虚拟平扫的衰减值一致性。各器官相关性极佳(p<0.0001),但配对t检验显示部分结构差异显著,且个体内VUE与TNC差值>10 HU的病例比例达13%~42%;VUE-静脉期总体更接近TNC。

要点

  • VUE与TNC各器官衰减值相关性极佳(p<0.0001),但个体内差值>10 HU的病例比例达13%~42%。
  • VUE-静脉期较VUE-动脉期总体更接近TNC衰减值。
  • 完全以VUE替代TNC前,可能需进一步优化后处理算法。

核心结论

  • VUE与TNC衰减值相关性极佳且平均差值可忽略(椎旁肌-5.94 HU至主动脉6.2 HU),但个体内分析显示相当数量病例VUE与TNC差值>10 HU。
  • VUE-静脉期总体更接近TNC值。
  • 在完全以VUE替代TNC图像之前,可能需进一步优化后处理算法。

研究设计与样本

  • 回顾性研究,纳入2014年8月至2015年5月间行腹盆三相CT的患者94例(肝脏肿块方案47例、胰腺肿块方案47例)。
  • 真实平扫采用常规单能模式120 kVp;动脉晚期及静脉期增强采用快速千伏切换双能CT技术。
  • 基于动脉期和静脉期系列分别生成VUE-art和VUE-ven图像,记录肝脏、胰腺、双肾、肾上腺、肌肉、皮下脂肪、主动脉、下腔静脉和主门静脉的TNC与VUE衰减值。
  • 采用单变量线性回归和Student双尾配对t检验比较衰减值。

关键影像表现/征象

  • TNC、VUE-art与VUE-ven衰减值在所有器官间相关性极佳(p<0.0001)。
  • 配对t检验显示TNC与VUE-art在双肾、右侧肾上腺、椎旁肌和主动脉衰减值上差异显著。
  • TNC与VUE-ven在左肾衰减值上差异显著。
  • 个体内VUE与TNC差值>10 HU的病例比例范围为13%(右肾)至42%(右肾上腺)。
  • TNC与VUE平均差值范围为椎旁肌-5.94 HU至主动脉6.2 HU。

诊断标准与关键数值

  • 以TNC衰减值为参考,VUE-art与VUE-ven各器官衰减值相关性极佳(p<0.0001)。
  • 个体内VUE与TNC差值>10 HU的病例比例:右肾13%、右肾上腺42%。
  • TNC与VUE平均差值:椎旁肌-5.94 HU、主动脉6.2 HU。

临床应用与阅片要点

  • VUE-静脉期较VUE-动脉期总体更接近TNC衰减值,阅片时可优先参考VUE-ven图像。
  • 双肾、右侧肾上腺、椎旁肌和主动脉的VUE-art与TNC差异显著,左肾的VUE-ven与TNC差异显著,报告时需注意。
  • 个体内VUE与TNC差值>10 HU的病例比例可达13%~42%,完全以VUE替代TNC前可能需进一步优化后处理算法。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: To assess the agreement between the true non-contrast (TNC) attenuation values of intra-abdominal structures and attenuation values obtained on virtual-unenhanced (VUE) images based on rapid kVp-switching dual-energy CT. The effects of contrast phase and patient characteristics (e.g., BMI, hematocrit, hemoglobin content) on VUE values were also investigated.

METHODS: Ninety four patients who underwent triphasic abdominal CT (liver mass protocol, n = 47; pancreas mass protocol, n = 47) between August 2014 and May 2015 were retrospectively reviewed. Unenhanced series was performed using conventional single-energy mode at 120 kVp. Late arterial and venous phase post-contrast series were obtained utilizing rapid kVp-switching dual-energy CT technique. VUE images were processed off of arterial (VUE-art) and venous (VUE-ven) phase series. Attenuation values of liver, pancreas, kidneys, adrenal glands, muscle, subcutaneous fat, aorta, IVC, and main portal vein were recorded on TNC and VUE sets of images. Attenuation values were compared using univariate linear regression and Student two-tailed paired t test.

RESULTS: There was excellent correlation between TNC, VUE-art, and VUE-ven attenuation values across all organs (p < 0.0001). Paired Student t test, however, showed significant difference between TNC and VUE-art attenuation of kidneys, right adrenal gland, paraspinal muscle, and aorta. There was also significant difference between TNC and VUE-ven attenuation of left kidney. Percentage of cases which had >10 HU difference between VUE and TNC for an individual was calculated which ranged between 13% (right kidney) and 42% (right adrenal gland).

CONCLUSION: Although the correlation between VUE and TNC attenuation values was excellent and mean difference between TNC and VUE attenuation values was negligible (ranging between -5.94 HU for paraspinal muscles to 6.2 HU in aorta), intra-patient analysis showed a considerable number of cases which had >10 HU difference between VUE and TNC. VUE-ven generally offered a better approximation of TNC values. Further optimization of post-processing algorithms might be necessary before complete replacement of TNC with VUE images.

来源:Abdom Radiol (NY),查看原文。版权归原作者所有。

引用

  1. Borhani AA, Kulzer M, Iranpour N, et al. Comparison of true unenhanced and virtual unenhanced (VUE) attenuation values in abdominopelvic single-source rapid kilovoltage-switching spectral CT. Abdom Radiol (NY). 2017. PMID: 27864600. doi:10.1007/s00261-016-0991-5

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。