血管造影C臂CT系统测量肝实质血容量的可重复性
在五只成年猪中,于肝左叶部分单纯栓塞前后行锥形束CT灌注成像,采用6秒延迟、双扫描8秒旋转方案,间隔10分钟采集三次。基线及栓塞后实质血容量图的平均变异系数分别为7%和25%,组内相关系数0.89,栓塞后血容量降低(P=0.007)。
要点
- 基线实质血容量图的平均变异系数为7%(范围2%-16%)。
- 栓塞后实质血容量图的平均变异系数为25%(范围13%-34%)。
- 实质血容量测量组内相关系数为0.89,显示高可重复性。
核心结论
- 使用锥形束CT在肝栓塞前后行术中三维实质血容量图采集具有高可重复性,在局部治疗中获取术中实质血容量图的应用前景良好。
研究设计与样本
- 动物实验(猪模型),5只成年猪,肝左叶部分单纯栓塞前后行锥形束CT成像,动脉内注射碘对比剂,6秒X线延迟,双扫描8秒旋转成像方案,每只动物栓塞前后各采集三次、间隔10分钟,专用软件后处理生成三维实质血容量图。
关键影像表现/征象
- 锥形束CT实质血容量图显示基线平均变异系数为7%(范围2%-16%)。
- 栓塞后实质血容量图平均变异系数为25%(范围13%-34%)。
- 实质血容量测量组内相关系数为0.89,显示高可重复性。
- 栓塞后显示可测量的实质血容量降低(P=0.007)。
诊断标准与关键数值
- 在每张锥形束CT三维实质血容量图的右叶和左叶两个视图上分别放置两个感兴趣区进行测量。
- 以变异系数和组内相关系数评估每只动物内扫描间实质血容量的变异。
- 采用Wilcoxon符号秩检验检验术后实质血容量降低。
临床应用与阅片要点
- 术中锥形束CT三维实质血容量图在肝栓塞前后采集具有高可重复性,可能有助于识别治疗终点或预测治疗结果。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
RATIONALE AND OBJECTIVES: Intra-procedural measurement of hepatic perfusion following liver embolization continues to be a challenge. Blood volume imaging before and after interventional procedures would allow identifying the treatment end point or even allow predicting treatment outcome. Recent liver oncology studies showed the feasibility of parenchymal blood volume (PBV) imaging using an angiographic C-arm system. This study was done to evaluate the reproducibility of PBV measurements using cone beam computed tomography (CBCT) before and after embolization of the liver in a swine model.
MATERIALS AND METHODS: CBCT imaging was performed before and after partial bland embolization of the left lobe of the liver in five adult pigs. Intra-arterial injection of iodinated contrast with a 6-second x-ray delay was used with a two-sweep 8-second rotation imaging protocol. Three acquisitions, each separated by 10 minutes to allow for contrast clearance, were obtained before and after embolization in each animal. Post-processing was carried out using dedicated software to generate three-dimensional (3D) PBV maps. Two region-of-interest measurements were placed on two views within the right and left lobe on each CBCT 3D PBV map. Variation in PBV for scans acquired within each animal was determined by the coefficient of variation and intraclass correlation. A Wilcoxon signed-rank test was used to test post-procedure reduction in PBV.
RESULTS: The CBCT PBV maps showed mean coefficients of variation of 7% (range: 2%-16%) and 25% (range: 13%-34%) for baseline and embolized PBV maps, respectively. The intraclass correlation for PBV measurements was 0.89, demonstrating high reproducibility, with measurable reduction in PBV displayed after embolization (P = 0.007).
CONCLUSIONS: Intra-procedural acquisition of 3D PBV maps before and after liver embolization using CBCT is highly reproducible and shows promising application for obtaining intra-procedural PBV maps during locoregional therapy.
引用
- Mueller K, Fahrig R, Manhart M, et al. Reproducibility of Parenchymal Blood Volume Measurements Using an Angiographic C-arm CT System. Acad Radiol. 2016. PMID: 27745815. doi:10.1016/j.acra.2016.08.001
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。