心脏运动对第三代双源CT支架管腔可视性的影响:基于搏动心脏模型的研究

Influence of cardiac motion on stent lumen visualization in third generation dual-source CT employing a pulsatile heart model.
PubMed 文献 · Br J Radiol · 2016-11-29

本研究在搏动心脏模型上植入13枚直径3.0 mm的冠状动脉支架,评估第三代双源CT(SOMATOM Force)在静止与运动状态下支架管腔可见度;运动状态支架管腔可见度较静止状态显著降低(60.8±4.4% vs 65.6±5.7%,p<0.001),但该 scanner 在心脏运动影响下仍能实现较高的支架管腔可见度。

要点

  • 运动状态支架管腔可见度60.8±4.4%,较静止状态65.6±5.7%显著降低(p<0.001)。
  • 该第三代双源CT在心脏运动影响下仍能实现较高的支架管腔可见度。
  • 结果是否可转化至临床情境尚需进一步患者研究评估。

核心结论

  • 静止与运动状态支架管腔可见度差异有统计学意义,但该第三代双源CT在心脏运动影响下仍能实现较高的支架管腔可见度,能否转化至临床情境尚需进一步患者研究评估。

研究设计与样本

  • 体外模型研究,在搏动心脏模型(CoroSim®)的充盈对比剂的塑料管中植入13枚直径3.0 mm的冠状动脉支架,使用第三代双源CT(SOMATOM Force)扫描,支架方向平行于扫描仪z轴,采用针对血管成像优化的中等锐利卷积核进行评价。

诊断标准与关键数值

  • 支架管腔可见度以百分比表示:静止状态支架为65.6±5.7%,运动状态支架为60.8±4.4%。

局限性

  • 原文结论指出,该结果能否转化至临床情境尚需进一步患者研究评估。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: To detect an in-stent restenosis, an invasive coronary angiography is commonly performed. Owing to the risk associated with this procedure, a non-invasive method to detect or exclude an in-stent restenosis is desirable. The purpose of this study was to evaluate the influence of cardiac motion on stent lumen visibility in a third-generation dual-source CT scanner (SOMATOM Force; Siemens Healthcare, Forchheim, Germany), employing a pulsatile heart model (CoroSim®; Mecora, Aachen, Germany).

METHODS: 13 coronary stents with a diameter of 3.0 mm were implanted in plastic tubes filled with a contrast medium and then fixed onto the pulsatile phantom heart model. The scans were performed while the heart model mimicked the heartbeat. Coronary stents were scanned in an orientation parallel to the scanner z-axis. The evaluation of the stents was performed by employing a medium sharp convolution kernel optimized for vascular imaging.

RESULTS: The mean visible stent lumen was reduced from 65.6 ± 5.7% for the stents at rest to 60.8 ± 4.4% for the stents in motion (p-value: <0.001).

CONCLUSION: While the difference in lumen visibility between stents in motion and at rest was significant, the use of this third-generation dual-source CT scanner enabled a high stent lumen visibility under the influence of cardiac motion. Whether this translates into a clinical setting has to be evaluated in further patient studies. Advances in knowledge: The employed modern CT scanner enables a high stent lumen visibility even under the influence of cardiac motion, which is important to detect or exclude an in-stent restenosis.

来源:Br J Radiol,查看原文。版权归原作者所有。

引用

  1. Petri N, Gassenmaier T, Allmendinger T, et al. Influence of cardiac motion on stent lumen visualization in third generation dual-source CT employing a pulsatile heart model. Br J Radiol. 2016. PMID: 27897030. doi:10.1259/bjr.20160616

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。