腹部双能CT血管成像中标准和先进虚拟单能成像窗设置的优化
研究纳入45例患者行第三代双源双能CT腹部血管成像,重建标准线性融合(M_0.6)、70 keV传统虚拟单能(M70)和40 keV先进噪声优化虚拟单能(M40+)三组数据,由两名盲法观察者确定最佳窗设置并与主动脉衰减相关得到优化窗设置。结果显示M70和M40+的最佳窗宽/窗位分别为880/280和1410/450,回归推断优化窗宽/窗位分别为850/270和1350/430;使用优化窗设置时M40+与标准M_0.6的血管测量无显著差异。
要点
- M70的最佳窗宽/窗位为880/280,M40+为1410/450。
- 回归推断的优化窗宽/窗位:M70为850/270,M40+为1350/430。
- M40+使用优化窗设置时血管测量与标准M_0.6无显著差异(p≥0.16)。
核心结论
- 为优化传统M70和先进M40+虚拟单能成像,调整窗宽/窗位是必要的;结果提示M70可采用850/270、M40+可采用1350/430的窗设置。
研究设计与样本
- 研究纳入45例患者,行第三代双源双能CT腹部血管成像(90/150 kV,180/90参考mAs),重建M_0.6、M70和M40+三组数据,由两名盲法观察者评估最佳窗设置并与主动脉衰减相关得到优化窗设置,采用重复测量方差分析比较窗值、图像质量和血管测量。
诊断标准与关键数值
- M70的最佳窗宽/窗位为880/280,M40+的最佳窗宽/窗位为1410/450。
- 回归分析推断M70的优化窗宽/窗位为850/270,M40+为1350/430。
- M40+的最佳与优化窗设置之间窗宽和窗位均有显著差异,M70与M40+之间在最佳和优化窗设置下窗宽和窗位均有显著差异。
- 使用优化窗设置时M40+与标准M_0.6的血管测量无显著差异(p≥0.16),使用最佳窗设置时M40+与M_0.6的血管测量有显著差异(p≤0.04)。
临床应用与阅片要点
- 作者建议传统M70虚拟单能成像采用850/270的窗宽/窗位设置。
- 作者建议先进M40+虚拟单能成像采用1350/430的窗宽/窗位设置。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: To determine the optimal window setting for displaying virtual monoenergetic reconstructions of third generation dual-source, dual-energy CT (DECT) angiography of the abdomen.
METHODS: Forty-five patients were evaluated with DECT angiography (90/150 kV, 180/90 ref. mAs). Three datasets were reconstructed: standard linear blending (M_0.6), 70 keV traditional virtual monoenergetic (M70), and 40 keV advanced noise-optimized virtual monoenergetic (M40+). The best window setting (width and level, W/L) was assessed by two blinded observers and was correlated with aortic attenuation to obtain the Optimized W/L setting (O-W/L). Subjective image quality was assessed, and vessel diameters were measured to determine any possible influences between different W/L settings. Repeated measures of variance were used to evaluate comparison of W/L values, image quality, and vessel sizing between M_0.6, M70, and M40+.
RESULTS: The Best W/L (B-W/L) for M70 and M40+ was 880/280 and 1410/450, respectively. Results from regression analysis inferred an O-W/L of 850/270 for M70 and 1350/430 for M40+. Significant differences for W and L were found between the Best and the Optimized W/L for M40+, and between M70 and M40+ for both the Best and Optimized W/L. No significant differences for vessel measurements were found using the O-W/L for M40+ compared to the standard M_0.6 (p ≥ 0.16), and significant differences were observed when using the B-W/L with M40+ compared to M_0.6 (p ≤ 0.04).
CONCLUSION: In order to optimize virtual monoenergetic imaging with both traditional M70 and advanced M40+, adjusting the W/L settings is necessary. Our results suggest a W/L setting of 850/270 for M70 and 1350/430 for M40+.