心脏SPECT/CT中的辐射剂量:SSDE与有效剂量的估算

Radiation dose in cardiac SPECT/CT: An estimation of SSDE and effective dose.
PubMed 文献 · Eur J Radiol · 2016-10-21

研究纳入509例心肌灌注SPECT/CT患者,基于AAPM指南和Impact-Dose软件估算SSDE、有效剂量及器官剂量。结果显示CTDIvol均值1.34 mGy±0.19、SSDE均值1.7 mGy±0.16,发射、CT及总有效剂量分别为11.5±1.4、0.49±0.11和12.67±1.73 mSv,作者认为剂量参数很低且低于参考水平。

要点

  • CTDIvol均值1.34 mGy±0.19,SSDE均值1.7 mGy±0.16,CT部分有效剂量仅0.49±0.11 mSv。
  • 发射剂量与患者体重强相关,与有效直径相关性较低。
  • CTDIvol强依赖患者体型,而SSDE不依赖患者体型。

核心结论

  • 本研究报告的CTDIvol、DLP、SSDE及有效剂量值很低且低于参考水平,提示SPECT/CT定位与衰减校正中采用适当的CT采集参数对患者有益并能降低癌症风险。

研究设计与样本

  • 纳入509例心肌灌注SPECT/CT患者,使用AAPM指南和Impact-Dose软件计算SSDE、有效剂量和器官剂量,采用R和SPSS软件进行Spearman-Pearson相关检验和线性回归分析。

诊断标准与关键数值

  • CTDIvol均值1.34 mGy±0.19,SSDE均值1.7 mGy±0.16。
  • 发射有效剂量11.5±1.4 mSv,CT有效剂量0.49±0.11 mSv,总有效剂量12.67±1.73 mSv。
  • 体型与器官剂量相关性显著,但甲状腺和红骨髓除外。

临床应用与阅片要点

  • 作者认为SPECT/CT定位与衰减校正中采用适当的CT采集参数对患者非常有益并能降低癌症风险。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

AIMS: The dose levels for Computed Tomography (CT) localization and attenuation correction of Single Photon Emission Computed Tomography (SPECT) are limited and reported as Volume Computed Tomography Dose Index (CTDIvol) and Dose-Length Product (DLP). This work presents CT dose estimation from Cardiac SPECT/CT based on new American Association of Physicists in Medicine (AAPM) Size Specific Dose Estimation (SSDE) parameter, effective dose, organ doses and also emission dose from nuclear issue.

MATERIAL AND METHODS: Myocardial perfusion SPECT/CT for 509 patients was included in the study. SSDE, effective dose and organ dose were calculated using AAPM guideline and Impact-Dose software. Data were analyzed using R and SPSS statistical software. Spearman-Pearson correlation test and linear regression models were used for finding correlations and relationships among parameters.

RESULTS: The mean CTDIvol was 1.34 mGy±0.19 and the mean SSDE was 1.7 mGy±0.16. The mean±SD of effective dose from emission, CT and total dose were 11.5±1.4, 0.49±0.11 and 12.67±1.73 (mSv) respectively. The mean±SD of effective dose from emission, CT and total dose were 11.5±1.4, 0.49±0.11 and 12.67±1.73 (mSv) respectively. The spearman test showed that correlation between body size and organ doses is significant except thyroid and red bone marrow. CTDIvol was strongly dependent on patient size, but SSDE was not. Emission dose was strongly dependent on patient weight, but its dependency was lower to effective diameter.

CONCLUSION: The dose parameters including CTDIvol, DLP, SSDE, effective dose values reported here are very low and below the reference level. This data suggest that appropriate CT acquisition parameters in SPECT/CT localization and attenuation correction are very beneficial for patients and lowering cancer risks.

来源:Eur J Radiol,查看原文。版权归原作者所有。

引用

  1. Abdollahi H, Shiri I, Salimi Y, et al. Radiation dose in cardiac SPECT/CT: An estimation of SSDE and effective dose. Eur J Radiol. 2016. PMID: 27842674. doi:10.1016/j.ejrad.2016.10.021

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。