文献速递胸部CT

一种评估肺癌CT筛查中肺结节计算机辅助检测性能的方法:结节大小与密度的检测限

A method for evaluating the performance of computer-aided detection of pulmonary nodules in lung cancer CT screening: detection limit for nodule size and density.
PubMed 文献 · Br J Radiol · 2017-01-03

研究提出用虚拟结节评估低剂量CT肺癌筛查中计算机辅助检测(CAD)肺结节的性能。虚拟结节基于机构所用CT系统的空间分辨率生成并融合入该机构临床肺图像,验证显示48个虚拟结节中46个与人工结节CAD结果一致(kappa系数=0.913),真实结节与对应虚拟结节CAD结果相同。测试的CAD系统检测限为:直径5 mm的周围型肺结节在结节与背景密度差ΔCT>220 HU时可被检出。

要点

  • 虚拟结节可替代人工结节用于CAD性能评估,48个中46个结果一致(kappa=0.913)。
  • 直径5 mm的周围型肺结节需ΔCT>220 HU时才能被该CAD系统检出。
  • 虚拟结节可体现特定机构的扫描/重建条件,用于评估该机构的CAD性能。

核心结论

  • 虚拟结节能有效评估特定机构扫描/重建条件下的CAD性能,并可指导结节大小/密度检测限的评估策略。

研究设计与样本

  • 方法学研究:基于机构所用CT系统测得的空间分辨率生成虚拟结节,融合入该机构临床肺图像;先以体模中人工结节验证虚拟结节,再比较真实结节(n=6)与对应虚拟结节CAD结果,最后将不同大小和密度差(ΔCT)的虚拟结节插入临床图像(n=10)进行CAD分析。

诊断标准与关键数值

  • 验证研究中48个虚拟结节有46个与人工结节CAD结果相同(kappa系数=0.913);真实结节与对应虚拟结节CAD结果相同;检测限:直径5 mm的周围型肺结节在ΔCT>220 HU时可被检出。

临床应用与阅片要点

  • 作者提出虚拟结节可作为评估特定机构CAD性能的有效手段,并认为指导结节大小/密度检测限的方法可能是有用的评估策略。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: We propose the application of virtual nodules to evaluate the performance of computer-aided detection (CAD) of lung nodules in cancer screening using low-dose CT.

METHODS: The virtual nodules were generated based on the spatial resolution measured for a CT system used in an institution providing cancer screening and were fused into clinical lung images obtained at that institution, allowing site specificity. First, we validated virtual nodules as an alternative to artificial nodules inserted into a phantom. In addition, we compared the results of CAD analysis between the real nodules (n = 6) and the corresponding virtual nodules. Subsequently, virtual nodules of various sizes and contrasts between nodule density and background density (ΔCT) were inserted into clinical images (n = 10) and submitted for CAD analysis.

RESULTS: In the validation study, 46 of 48 virtual nodules had the same CAD results as artificial nodules (kappa coefficient = 0.913). Real nodules and the corresponding virtual nodules showed the same CAD results. The detection limits of the tested CAD system were determined in terms of size and density of peripheral lung nodules; we demonstrated that a nodule with a 5-mm diameter was detected when the nodule had a ΔCT > 220 HU.

CONCLUSION: Virtual nodules are effective in evaluating CAD performance using site-specific scan/reconstruction conditions. Advances in knowledge: Virtual nodules can be an effective means of evaluating site-specific CAD performance. The methodology for guiding the detection limit for nodule size/density might be a useful evaluation strategy.

来源:Br J Radiol,查看原文。版权归原作者所有。

引用

  1. Kobayashi H, Ohkubo M, Narita A, et al. A method for evaluating the performance of computer-aided detection of pulmonary nodules in lung cancer CT screening: detection limit for nodule size and density. Br J Radiol. 2017. PMID: 27897029. doi:10.1259/bjr.20160313

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。