单次采集双成像:改良DIXON联合FRACTURE序列同时显示腰椎骨质与神经结构
骨质评估中3D-mDIXON同相位图像CNR显著高于常规FRACTURE,神经评估中其水图像CNR与3D-ProSet无显著差异,但整体偏好和运动伪影评分显著更优,提示单次采集可在保持图像质量的同时提高效率并便于评估空间关系。
要点
- 3D-mDIXON单次采集可同时获得骨质(同相位)和神经(水)图像,便于评估两者空间关系。
- 骨质评估中3D-mDIXON同相位图像CNR显著高于常规FRACTURE。
- 神经评估中3D-mDIXON水图像CNR与3D-ProSet无显著差异,但整体偏好和运动伪影评分显著更优。
核心结论
- 3D-mDIXON技术可在单次采集同时显示骨质与神经结构并保持足够图像质量,可能提高成像效率并便于评估两者空间关系。
研究设计与样本
- 将mDIXON辅助FRACTURE(3D-mDIXON)获得的骨质(同相位)和神经(水)图像分别与常规FRACTURE和3D-FFE-ProSet(3D-ProSet)比较。
- 图像性能采用对比噪声比(CNR)和视觉评估评价,CNR和视觉评估评分用Wilcoxon符号秩检验比较,观察者间一致性用ICC[A,2]评估。
诊断标准与关键数值
- 骨质评估:3D-mDIXON(同相位)的CNR显著高于FRACTURE(p < 0.01)。
- 神经评估:3D-mDIXON(水)与3D-ProSet的CNR无显著差异。
- 3D-mDIXON的整体偏好评分(p < 0.01)和运动伪影评分(p < 0.01)显著高于3D-ProSet。
临床应用与阅片要点
- 可利用3D-mDIXON单次采集同时获得骨质和神经图像,在腰椎MRI中评估两者空间关系。
- 骨质评估可参考同相位图像,神经评估可参考水图像,注意其CNR与常规序列的差异。
局限性
- 单中心前瞻性可行性研究,样本为健康志愿者,未纳入患者。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: Simultaneous evaluation of osseous and neural structures is important in assessing lumbar spine pathologies. However, conventional magnetic resonance imaging (MRI) sequences cannot depict both structures simultaneously, and multiple sequences may prolong scanning time and cause spatial misregistration. This preliminary feasibility study investigated whether modified Dixon (mDIXON, Philips Healthcare)-assisted fast field echo resembling computed tomography using restricted echo-spacing (FRACTURE, Philips Healthcare) enabled simultaneous depiction of both structures from a single acquisition.
MATERIALS AND METHODS: In this single-center prospective study, healthy volunteers underwent lumbar spine MRI using a 1.5-T MRI system. Bone (in-phase) and nerve (water) images obtained using mDIXON-assisted FRACTURE (3D-mDIXON) were compared with conventional FRACTURE and 3D-FFE-ProSet (3D-ProSet), respectively. Image performance was evaluated using the contrast-to-noise ratio (CNR) and visual assessment. The Wilcoxon signed-rank test was used to compare CNR and visual assessment scores. Interobserver agreement was assessed according to the intraclass correlation coefficient (ICC[A,2]).
RESULTS: Twenty healthy volunteers (mean age, 48.2 ± 9.6 years) were included. For bone evaluation, the CNR was significantly higher with 3D-mDIXON (in-phase) than with FRACTURE (p < 0.01). For nerve evaluation, CNR did not significantly differ between 3D-mDIXON (water) and 3D-ProSet. However, the overall preference score (p < 0.01) and motion artifact score (p < 0.01) were significantly higher for 3D-mDIXON than for 3D-ProSet.
CONCLUSIONS: The 3D-mDIXON technique enables the simultaneous depiction of both structures from a single acquisition while maintaining adequate image quality. This approach may improve imaging efficiency and facilitate assessment of their spatial relationship.
引用
- Taguchi A, Jomoto W, Takatsu Y, et al. One-shot dual imaging: simultaneous visualization of bone and nerve structures using modified DIXON and fast field echo resembling computed tomography with restricted echo-spacing magnetic resonance imaging. Eur J Radiol. 2026. PMID: 42913347. doi:10.1016/j.ejrad.2026.113287
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。