文献速递心血管儿科影像MR

建立1.5T非对比增强4D flow在先天性心脏病中的可靠性

Establishing the reliability of non-contrast 4D flow in congenital heart disease at 1.5T.
PubMed 文献 · Pediatr Radiol · 2026-10-13

前瞻性纳入30例8-20岁疑似或确诊先天性心脏病患者,在1.5T行心脏MRI采集非对比增强4D flow与2D相位对比血流。非对比增强4D flow与2D PC在净流量、峰值流速和反流分数上相关性好,但净流量低估约10.1%、峰值流速高估约18.9%,心室每搏量也略有低估;两者内部一致性良好,SNR/CNR无差异。

要点

  • 非对比增强4D flow与2D PC净流量、峰值流速和反流分数相关性好。
  • 4D flow净流量较2D PC低约10.1%,峰值流速高约18.9%。
  • 4D flow低估心室每搏量,升主动脉较左室每搏量低18.1%。

局限性

  • 样本量较小(30例)。
  • 单中心前瞻性研究。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

BACKGROUND: 4D flow has been shown to correlate well with 2D phase-contrast (2D PC) flow measurements, although most of the current literature is based on post-contrast imaging.

OBJECTIVE: Given recent interest in non-contrast techniques, we hypothesized that non-contrast 4D flow (NC 4D Flow) would correlate well with 2D PC and ventricular stroke volumes.

METHODS AND MATERIALS: A prospective cohort of pediatric and young adult patients with suspected/confirmed congenital heart disease (CHD) underwent cardiac MRI at 1.5T with acquisition of NC 4D Flow and 2D PC. Pearson correlation, linear mixed-effect correlations, and Bland-Altman comparisons were performed to compare these methods with secondary analyses of internal consistency and image quality/noise.

RESULTS: In our cohort of 30 patients (age, 8-20 years), NC 4D Flow correlated strongly with 2D PC for net flow (r=0.76), peak velocity (r=0.9, P<0.001), and regurgitant fraction (r=0.75, P<0.001). By Bland-Altman comparison, NC 4D Flow was 10.1% lower for net flows and 18.9% higher for peak velocity compared to 2D PC. NC 4D Flow also slightly underestimated ventricular stroke volumes (18.1% lower for ascending aorta vs. left ventricular stroke volume and 13.9% lower for main pulmonary artery vs. right ventricular stroke volume). Both datasets were internally consistent; 2D PC had a higher relative signal intensity, but no difference in signal-to-noise or contrast-to-noise ratios (SNR/CNR).

CONCLUSION: Non-contrast 4D flow correlates well with 2D PC in CHD with a slight underestimation in flows and overestimation in peak velocities, but good internal consistency and comparable SNR/CNR.

来源:Pediatr Radiol,查看原文。版权归原作者所有。