多发性骨髓瘤患者双源光子计数CT全身扫描中股骨骨髓衰减的重建核依赖

Reconstruction-kernel Dependence of Femoral Bone Marrow Attenuation in Whole-body Dual-source Photon-counting CT of Patients with Multiple Myeloma.
PubMed 文献 · Acad Radiol · 2026-09-01

回顾性纳入103例多发性骨髓瘤患者,行非增强全身光子计数CT(超高清双源模式,90 kV/Sn 150 kV),从70 keV虚拟单能图像重建Br32–Br68及Qr32–Qr68核。以Qr40为参考,Br32、Br36、Br40的股骨骨髓衰减显著更低(均p<0.001),而Br48–Br68及所有定量核间无显著差异;图像噪声随核锐度增加而升高。结论:股骨骨髓衰减依赖重建核,定量评估需严格标准化重建。

要点

  • 股骨骨髓CT衰减值随重建核变化,Br32–Br40较Qr40显著偏低,不可视为重建不变。
  • 定量骨髓评估应以Qr40等定量核为参考,并固定重建方案以保证随访可比性。
  • 图像噪声随核锐度增加而升高,选择核时需权衡噪声与骨髓衰减准确性。

关键影像表现/征象

  • 以定量核Qr40为参考(中位-66.9 HU,IQR -82.0至-45.5 HU),Br32、Br36、Br40的股骨骨髓衰减显著更低(均p<0.001)。
  • 更锐利的标准体部核Br48至Br68之间骨髓衰减无显著差异(均p>0.999)。
  • 所有定量核之间骨髓衰减无显著差异,与锐度无关(均p≥0.108)。
  • 肌肉和脂肪衰减在各重建间保持稳定(均p≥0.077)。
  • 体部核和定量核的图像噪声均随核锐度增加而逐渐升高(均p≤0.007)。

诊断标准与关键数值

  • 以定量核Qr40为参考标准,股骨骨髓衰减中位-66.9 HU(IQR -82.0至-45.5 HU)。
  • Br32骨髓衰减中位-95.9 HU(IQR -113.1至-73.4 HU),Br36中位-96.0 HU(IQR -111.9至-73.0 HU),Br40中位-89.5 HU(IQR -102.7至-65.1 HU),均显著低于Qr40(p<0.001)。

临床应用与阅片要点

  • 多发性骨髓瘤骨髓定量评估应固定使用定量核(如Qr40)并标准化重建方案,避免随访中更换核导致衰减差异误判为骨髓成分变化。
  • 报告骨髓CT衰减值时需注明所用重建核,不同核间的衰减值不可直接比较。
  • 选择重建核时需考虑噪声随锐度增加而升高,平衡图像质量与测量稳定性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

RATIONALE AND OBJECTIVES: Reconstruction kernels have various effects on image characteristics. This study aims to investigate how kernel selection influences the femoral bone marrow attenuation in patients with multiple myeloma.

MATERIALS AND METHODS: This retrospective study included 103 consecutive patients who underwent noncontrast whole-body photon-counting computed tomography (PCCT) for assessment of multiple myeloma. All examinations were performed in ultrahigh-resolution dual-source mode (96 × 0.2 mm) at 90 kV/Sn 150 kV. For each examination, standard body regular (Br) kernels from Br32 to Br68 and the corresponding quantitative kernels (i.e., Qr32-Qr68) were reconstructed from 70 keV virtual monoenergetic images. Regions of interest were placed bilaterally in the femoral bone marrow, the adjacent muscle, and subcutaneous fat. Marrow attenuation and image noise were compared across kernels.

RESULTS: Compared to the quantitative kernel Qr40 as reference standard (median -66.9 [interquartile range -82.0--45.5] hounsfield units [HU]), marrow attenuation was substantially lower with Br32 (-95.9 [-113.1--73.4] HU), Br36 (-96.0 [-111.9--73.0] HU), and Br40 (-89.5 [-102.7--65.1] HU) (all p < 0.001). Meanwhile, attenuation values did not differ significantly among the sharper standard body kernels Br48 to Br68 (all p > 0.999). No attenuation difference was ascertained among the quantitative kernels, regardless of sharpness (all p ≥ 0.108). Muscle and fat attenuation remained stable across reconstructions (all p ≥ 0.077). Image noise increased progressively with higher kernel sharpness in both body and quantitative kernels (all p ≤ 0.007).

CONCLUSION: Femoral bone marrow attenuation in dual-source PCCT is reconstruction-kernel dependent. Quantitative marrow assessment in patients with multiple myeloma, therefore, requires strict reconstruction standardization, and attenuation measurements should not be considered reconstruction-invariant. Otherwise, attenuation differences may reflect reconstruction choice rather than changes in marrow composition.