要点

  • 机器人辅助脑血管造影在效率、对比剂用量和辐射剂量上与手动无显著差异。
  • 机器人组18%需中转手动,常见原因为血管迂曲、导管不兼容或卡匣错误。
  • 操作经验积累与手术时间缩短、对比剂用量及辐射剂量降低独立相关,学习曲线约50例。

局限性

  • 单中心回顾性设计
  • 机器人组中转率较高(18%)可能影响意向性治疗分析结果
  • 学习曲线仅评估3名术者

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

BACKGROUND: Diagnostic cerebral angiography (DCA) exposes neurointerventionalists to substantial occupational radiation. Robotic-assisted cerebral angiography (RCA) offers operator-dose reduction and a foundation for telerobotic stroke care, but comparisons to manual DCA have only been evaluated in smaller feasibility cohorts.

METHODS: We performed a single-institution retrospective cohort study between September 2021 and May 2023, comparing 214 procedures using the CorPath GRX system against 636 manual DCA controls. Primary outcomes were procedural efficiency (room, procedure, and fluoroscopy time) and patient radiation dose (cumulative air kerma), with per-vessel normalization. Robotic cases were analyzed on an intention-to-treat basis. The operator learning curve was assessed among three neurointerventional primary operators at the attending level, each of whom performed at least 10 cases.

RESULTS: Robotic and manual angiography did not differ significantly in room time, procedure time, fluoroscopy time, contrast volume administered, or radiation dose. Per-vessel comparisons were similarly null (all p≥0.51). Findings were unchanged after adjustment for age, sex, number of vessels catheterized, access site, and operator. Intraoperative complications were rare (0% vs. 0.8%, p=0.34). Robotic cases required conversion to manual angiography in 18% of cases (39/214), most commonly due to tortuous anatomy, catheter incompatibility, or cassette error. Increasing procedural experience was independently associated with shorter procedural times, and lower contrast administered and patient radiation dose. The learning curve extended across ∼50 cases for outcomes such as room and procedure time.

CONCLUSIONS: RCA showed no statistically significant differences from manual DCA in terms of efficiency, contrast use, and safety, with a learning curve spanning around 50 cases.