PETRA MRA与传统MRA序列在颅内动脉瘤随访中的真实世界对比研究
结果显示PETRA的血管可见度优于TOF-MRA、与ceMRA相当,伪影所致信号丢失以TOF-MRA最重,但PETRA整体图像质量最低且采集时间较长,各序列对再灌注的检出无差异。
要点
- PETRA血管可见度优于TOF-MRA并与ceMRA相当,但整体图像质量最差。
- TOF-MRA伪影相关信号丢失和受损最严重,与ceMRA和PETRA差异显著。
- 各MRA序列对动脉瘤再灌注的检出无差异,PETRA可作为非对比增强随访选项。
关键影像表现/征象
- PETRA的血管可见度评分优于TOF-MRA,与ceMRA相当。
- TOF-MRA的伪影所致信号丢失和信号受损最严重,与ceMRA和PETRA差异有统计学意义。
- PETRA的整体图像质量评分最低。
- 各MRA序列对动脉瘤再灌注的检出无差异。
诊断标准与关键数值
- 采用3分制评估动脉瘤再灌注、图像质量及载瘤动脉和分支血管可见度。
- 伪影相关信号丢失和信号减低以毫米为单位测量。
- 采用Fleiss Kappa评估评分者间信度。
临床应用与阅片要点
- 对植入物相关伪影明显的动脉瘤随访,PETRA可提供优于TOF的血管可见度且无需对比剂。
- 需注意PETRA整体图像质量较低,阅片时应结合ceMRA综合判断。
- 各序列再灌注检出无差异,序列选择可综合考虑伪影、对比剂禁忌和扫描时间。
局限性
- PETRA采集时间较长,限制临床可行性和真实世界适用性。
- PETRA整体图像质量最低,可能影响细节评估。
- 需进一步研究缩短扫描时间并评估跨装置性能。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND AND PURPOSE: MRI follow-up of treated cerebral aneurysms is essential to detect remnants or reperfusion but contrast-enhanced MR-angiography (ceMRA) and TOF-MRA can be impaired by implant-related artifacts. Pointwise Encoding Time Reduction with Radial Acquisition MRA (PETRA) is a novel, non-contrast MRA technique with promising results for aneurysms treated with clipping or coiling. However, its performance in patients treated with other devices, including flow diverters and intrasaccular devices, remains unclear. This study aimed to compare PETRA with other MRA sequences with a particular focus on vessel visibility.
MATERIALS AND METHODS: Ninety-seven patients with 124 aneurysms treated with various endovascular devices or surgical clips were included. All patients underwent MRI including PETRA, TOF-MRA and ceMRA between September 2024 and February 2026. Three neuroradiologists independently rated aneurysm reperfusion, as well as image quality and the visibility of the parent artery and branching vessels, using 3-point scales. Artifact-related signal loss and signal reduction due to artifacts were measured in millimeters. Interrater reliability was determined using Fleiss Kappa.
RESULTS: Vessel visibility was rated superior with PETRA compared to TOF-MRA, and comparable to ceMRA. Artifact-induced signal loss and impairment were highest with TOF-MRA, differing significantly from ceMRA and PETRA. Overall image quality was lowest with PETRA. Reperfusion detection did not differ among the MRA sequences.
CONCLUSIONS: PETRA shows potential as a non-contrast MRA technique for follow-up of cerebral aneurysms, providing better vessel visibility than TOF and diagnostic performance comparable to ceMRA. However, its prolonged acquisition time and associated lower image quality currently limits clinical feasibility and real-world applicability. Further studies focusing on scan time reduction and cross-device performance are required.
引用
- Sondermann S, Schramm P, Keric N, et al. Comparison of Pointwise Encoding Time Reduction with Radial Acquisition (PETRA) MR-Angiography with conventional MR-Angiography-Sequences for Follow-Up of intracranial Aneurysms in a real-world cohort. AJNR Am J Neuroradiol. 2026. PMID: 42810827. doi:10.3174/ajnr.A9652
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。