视觉不可区分的声耦合垫用于双盲聚焦超声神经调控研究
本研究开发并验证了一对视觉和机械上不可区分的声耦合垫,分别用于高效超声传输和有效超声阻断,以实现真正的双盲LIFU神经调控研究。传输垫平均衰减为-0.41±0.53 dB,满足最小声损失的设计标准;非传输垫平均衰减为-48.61±4.33 dB,超过有效伪刺激所需的-40 dB阈值。两种垫在视觉和触觉上无法区分,且在反复清洁后保持机械完整性。
要点
- 非传输垫在焦点平面产生扩散波束,无清晰焦点位置。
- 传输垫平均衰减-0.41±0.53 dB,满足<5%压力振幅损失的设计标准。
- 非传输垫平均衰减-48.61±4.33 dB,超过-40 dB阈值。
关键影像表现/征象
- 非传输垫在焦点平面产生扩散波束,无清晰焦点位置。
- 传输垫平均衰减为-0.41±0.53 dB,满足最小声损失的设计标准。
- 非传输垫平均衰减为-48.61±4.33 dB,超过有效伪刺激所需的-40 dB阈值。
- Dragon Skin 10 NV基底声速为964.72 m/s,标准垫厚度下产生<2 mm轴向焦点偏移,焦点宽度无显著变化。
- 两种垫在视觉和触觉上无法区分,经验丰富的操作者或参与者均无法区分。
诊断标准与关键数值
- 传输垫设计标准:相对于自由水传播至目标焦点区域,压力振幅损失<5%。
- 非传输垫设计标准:在相同焦点平面衰减超声≥40 dB。
- 传输垫实测平均衰减-0.41±0.53 dB,满足设计标准。
- 非传输垫实测平均衰减-48.61±4.33 dB,超过-40 dB阈值。
临床应用与阅片要点
- 使用视觉和触觉不可区分的耦合垫可在LIFU神经调控研究中实现真正的双盲条件。
- 非传输垫产生扩散波束且无清晰焦点,可作为有效伪刺激。
- 传输垫声学损失小且焦点偏移<2 mm,适用于精确靶向。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND: Low intensity focused ultrasound (LIFU) is an emerging noninvasive neuromodulation modality with the ability to target deep brain structures with high spatial precision. Despite its promise, rigorous evaluation of its efficacy is limited by the absence of reliable, fully double-blind sham methodologies.
OBJECTIVE: To develop and validate a pair of visually and mechanically indistinguishable acoustic coupling pads that enable true double-blind LIFU neuromodulation studies by providing either efficient ultrasound transmission or robust ultrasound blocking without altering participant or operator experience.
METHODS: Two coupling pads were engineered: a transmitting pad designed to allow <5% pressure amplitude loss relative to free-water propagation to the intended focal region, and a non-transmitting pad designed to attenuate ultrasound by ≥40 dB in the same focal plane. Both pads used a Dragon Skin 10 NV silicone base and were identical in size, appearance, flexibility, and handling. The non-transmitting pad incorporated an encapsulated air-based blocking layer using a polyethylene foam insert. Acoustic performance was evaluated in a water tank using a 650 kHz BrainSonix transducer and a calibrated needle hydrophone including both waveform and spatial measurements. Sound speed of the silicone material was measured using pulse-echo techniques.
RESULTS: Twenty-three matched transmitting and non-transmitting pad pairs were fabricated and tested. Transmitting pads demonstrated a mean attenuation of -0.41 ± 0.53 dB, satisfying the design criterion of minimal acoustic loss. Non-transmitting pads demonstrated a mean attenuation of -48.61 ± 4.33 dB, exceeding the required -40 dB threshold for effective sham conditions. The Dragon Skin 10 NV substrate exhibited a sound speed of 964.72 m/s and produced <2 mm axial focal shift for standard pad thicknesses, with no measurable change in focal width. Non-transmitting pads produced a diffused beam profile in the focal plane, with no clear focal location. Both pad types were visually and tactually indistinguishable, could not be differentiated by experienced operators or participants, and maintained mechanical integrity after repeated cleaning.
CONCLUSION: These acoustically engineered coupling pads provide a practical and validated solution for achieving true single- and double-blind conditions in LIFU neuromodulation studies. By preserving identical sensory and procedural experiences while enabling precise control over ultrasound transmission, this approach addresses a critical methodological gap in human ultrasound neuromodulation research.
引用
- Schafer SF, Spivak NM, Bishay AAED, et al. Development of Visually Indistinguishable Acoustic Coupling Pads for Double-Blind Focused Ultrasound Neuromodulation Studies. Ultrasound Med Biol. 2026. PMID: 42595594. doi:10.1016/j.ultrasmedbio.2026.07.007
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。