深度学习增强与降噪用于低碘低辐射70kVp脑CTA的前瞻性研究
该前瞻性研究纳入47名健康志愿者,随机分为三组不同CTA协议,比较100kVp常规碘量与70kVp低碘/超低碘方案的图像质量。结果显示70kVp方案辐射剂量降低23.7%,动脉衰减增高但噪声增加;应用深度学习增强对比和降噪算法后,动脉衰减提高45.4%、噪声降低34.5%,CNR和主观图像质量显著优于常规方案。
要点
- 70kVp低碘脑CTA辐射剂量较100kVp降低23.7%,但图像噪声增加。
- 深度学习增强对比和降噪算法可使动脉衰减提高45.4%、噪声降低34.5%。
- 经深度学习处理后,70kVp方案的CNR和主观图像质量显著优于100kVp常规方案。
关键影像表现/征象
- 与A组相比,B组和C组辐射剂量降低23.7%。
- 滤波反投影重建下,B组动脉衰减为435.8±50.2 HU、C组为391.8±52.1 HU,均显著高于A组的321.1±47.4 HU(P<0.001)。
- 滤波反投影重建下,三组CNR无显著差异(A组19.9±4.7,B组20.3±3.8,C组18.4±4.6),因B组和C组噪声较高。
- 应用深度学习增强对比和降噪算法后,B组和C组动脉衰减提高45.4%、噪声降低34.5%,动脉衰减、CNR和主观图像质量均显著高于A组(P<0.001)。
诊断标准与关键数值
- 动脉衰减测量值:A组321.1±47.4 HU,B组435.8±50.2 HU,C组391.8±52.1 HU(滤波反投影重建)。
- CNR值:A组19.9±4.7,B组20.3±3.8,C组18.4±4.6(滤波反投影重建)。
- 深度学习增强对比和降噪算法使动脉衰减提高45.4%、噪声降低34.5%。
临床应用与阅片要点
- 70kVp低碘/超低碘脑CTA方案可降低辐射剂量,但需配合深度学习重建算法以抑制噪声。
- 深度学习增强对比和降噪算法能改善低碘低辐射CTA的动脉显示和图像质量,有望用于临床。
局限性
- 研究仅纳入健康志愿者,未在患者群体中验证。
- 样本量较小(共47例),分组后每组15-16例。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: To evaluate the feasibility of cerebral computed tomography angiography (CTA) obtained with reduced iodine and low radiation at 70 kVp and the effect of deep learning-based augmented contrast enhancement (DL-ACE) and denoising (DL-DN) algorithms on the CTA quality.
MATERIALS AND METHODS: In this prospective study, 47 healthy volunteers (male:female, 31:16; mean age ± standard deviation, 57.8 ± 10.9 years) were randomly assigned to one of three CTA protocols: Group A (n = 16; 100 kVp, 40 mL of 350 mgI/mL), Group B (n = 16; 70 kVp, 40 mL of 270 mgI/mL), and Group C (n = 15; 70 kVp, 28 mL of 270 mgI/mL [ultralow iodine]), with an injection rate of 2.5 mL/s for all. Images were reconstructed using filtered back projection (FBP), and images in Groups B and C were additionally reconstructed using DL-ACE and DL-DN. Arterial attenuation, image noise, contrast-to-noise ratio (CNR), and subjective image quality were compared among five image sets.
RESULTS: Compared with Group A, Groups B and C received 23.7% lower radiation doses. With FBP, arterial attenuation was significantly higher in Groups B (435.8 ± 50.2 Hounsfield units [HU]) and C (391.8 ± 52.1 HU) than in Group A (321.1 ± 47.4 HU) (P < 0.001), while CNR did not differ significantly (Group A, 19.9 ± 4.7; Group B, 20.3 ± 3.8; and Group C, 18.4 ± 4.6) due to higher image noise in Groups B and C. After applying DL-ACE and DL-DN in Groups B and C, arterial attenuation increased by 45.4% and image noise decreased by 34.5%, resulting in significantly higher arterial attenuation, CNR, and subjective image quality compared with Group A (P < 0.001).
CONCLUSION: Cerebral CTA at 70-kVp using ultralow iodine enhanced arterial attenuation but increased image noise compared with the 100-kVp CTA protocol. DL-ACE and DL-DN significantly increased arterial attenuation and reduced image noise, resulting in higher CNR and better subjective image quality.