CT自动检测肾结石的准确性与精确性:体模与临床研究

Accuracy and precision of automated kidney stone detection on CT.
PubMed 文献 · Abdom Radiol (NY) · 2026-05-20

本研究采用体模(120颗已知最大径1.4–9.9 mm、不同成分的肾结石)和45例患者CT,评估自研定量结石分析软件(qSAS)检测结石及测量最大径的准确性与精确性。qSAS在1、3、5 mm层厚的检测频率分别为100%、93%、86%;直径误差>1 mm的比例分别为5%、28%、64%。1 mm层厚时CT测量直径准确且不受结石成分影响;较厚层厚下尿酸结石倾向低估、钙及混合结石倾向高估。放射科医生测量相对无偏但精确性有限,直径误差>1 mm者占51%。

要点

  • 1 mm层厚CT联合自动软件检测结石频率100%,直径误差>1 mm仅5%。
  • 层厚增至3或5 mm时检测频率降至93%和86%,直径误差>1 mm比例升至28%和64%。
  • 较厚层厚下尿酸结石直径倾向低估,钙及混合结石倾向高估。
英文原文(摘要)· 点击展开

BACKGROUND: While computed tomography (CT) is the preferred imaging modality for kidney stone detection and measurement, quantifying the accuracy and precision of stone size metrics is needed, particularly with the growing use of automated image analysis tools to measure kidney stone size.

METHODS: A phantom study was conducted using 120 kidney stones of various known maximum diameters (ranging from 1.4 to 9.9 mm) and compositions (60 calcium, 30 uric acid, and 30 mixed calcium and uric acid). The stones were placed in an anthropomorphic phantom and CT images were reconstructed with slice thicknesses of 1, 3, or 5 mm and processed using an in-house quantitative Stone Analysis Software (qSAS) to detect each stone and determine its maximum diameter. Separately, we also assessed the agreement between qSAS and a radiologist in determining the maximum diameter of 45 kidney stones on CT images (1-mm slice) from patients.

RESULTS: Stone detection frequencies by qSAS for stones in phantoms were 100%, 93%, and 86% for 1-, 3-, and 5-mm CT slice thicknesses, respectively. The proportion of qSAS-detected stones with a diameter error > 1 mm was 5%, 28%, and 64% for 1-, 3-, and 5-mm CT slice thickness, respectively. The CT-measured diameters on 1-mm images were accurate regardless of stone composition. However, diameters with thicker slices trended toward underestimation in uric acid stones and overestimation in calcium and mixed stones. Stone diameter by a radiologist was relatively unbiased but had limited precision with a diameter error > 1 mm occurring in 51% compared to qSAS.

CONCLUSIONS: This study quantifies the improved precision with an automated assessment of stone diameter on CT as well as the bias and imprecision that remains across different slice thicknesses and stone compositions.

来源:Abdom Radiol (NY),查看原文。版权归原作者所有。