文献速递神经外周血管CT

高浓度对比剂结合低剂量扫描方案降低头颅CTA辐射剂量

Achieving lower radiation dose of cerebral computed tomography angiography: The role of high-concentration contrast medium.
PubMed 文献 · Eur J Radiol · 2026-05-17

前瞻性研究纳入150例行头颅CTA患者,随机分三组,均接受总碘量21.6g、80kV扫描。A组用碘美普尔400、IDR 2.0g/s、90mAs;B组用碘普罗胺370、IDR 1.48g/s、164mAs;C组用碘普罗胺370、IDR 2.0g/s、90mAs。结果显示A、C组颅内血管SNR和CNR与B组无显著差异,但辐射剂量较B组分别降低46.53%和45.62%,A组注射流率更低。

要点

  • 高碘浓度对比剂(碘美普尔400)可在较低注射流率下达到目标IDR,降低辐射剂量约46%。
  • 在80kV管电压下,采用高IDR(2.0g/s)联合低mAs(90mAs)方案不降低颅内血管SNR和CNR。
  • 低剂量方案下图像噪声增高,但CT衰减值也增高,需注意窗宽窗位调整。

关键影像表现/征象

  • A组和C组颅内血管CT衰减值及图像噪声均高于B组,但三组间SNR和CNR无统计学显著差异。
  • A组和C组辐射剂量较B组分别降低46.53%和45.62%。
  • A组注射流率低于C组,仍能达到目标IDR 2.0g/s。

诊断标准与关键数值

  • A组方案:碘美普尔400,IDR 2.0g/s,管电流时间乘积90mAs,管电压80kV,总碘量21.6g。
  • B组方案:碘普罗胺370,IDR 1.48g/s,管电流时间乘积164mAs,管电压80kV,总碘量21.6g。
  • C组方案:碘普罗胺370,IDR 2.0g/s,管电流时间乘积90mAs,管电压80kV,总碘量21.6g。

临床应用与阅片要点

  • 在头颅CTA中可优先考虑高浓度碘对比剂(如碘美普尔400)联合低mAs(90mAs)和80kV方案,以降低辐射剂量。
  • 使用低剂量方案时图像噪声增高,阅片时需适当调整窗宽窗位,注意评估小血管显示是否受影响。
  • 高IDR(2.0g/s)可通过较低注射流率实现,有助于减少注射相关风险。

局限性

  • 研究为单中心前瞻性设计,样本量150例,结论外推需谨慎。
  • 未报告诊断准确性(敏感度、特异度、AUC)等临床效能指标。
  • 仅评估了客观图像质量指标,未纳入主观图像质量评分。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: To investigate whether high-concentration iodine contrast medium (HICM) combined with low-dose radiation scanning protocols can further reduce the radiation dose of cerebral computed tomography angiography (CTA) scans without compromising image quality.

MATERIALS AND METHODS: The prospective study included 150 eligible patients who underwent cranial CTA examinations. The patients were randomly divided into three groups (A, B, and C). All patients received a total iodine dose of 21.6 g and underwent scanning at tube voltage of 80 kV. Each group included 50 patients. Different protocols were applied as below. Group A: Iomeprol 400, iodine delivery rate (IDR) 2.0 g/s, tube current-time product (TCTP) 90 mAs; Group B: Iopromide 370, IDR 1.48 g/s, TCTP 164 mAs; Group C: Iopromide 370, IDR 2.0 g/s, TCTP 90 mAs. Radiation doses (Computer Tomography Dose Index Volume, Dose-Length Product, Effective Dose) were recorded, and CT attenuation values, image noise, Signal-to-Noise Ratio (SNR), and Contrast-to-Noise Ratio (CNR) were measured.

RESULTS: There were no statistically significant differences in the SNR and CNR of intracranial vessels among groups A, B, and C. Group A and C showed higher CT attenuation values and image noise, but lower radiation dose (reduced by 46.53% and 45.62% compared with Group B). The injection flow rate was lower in Group A.

CONCLUSION: A higher IDR is effective in reducing radiation dose. Moreover, higher iodine concentration (iomeprol 400) enables a lower injection flow rate to achieve a target IDR. A 46% reduction in radiation dose can be achieved using iomeprol 400 combined with a low-dose protocol, without compromising image quality, at an optimized injection flow rate.

来源:Eur J Radiol,查看原文。版权归原作者所有。

引用

  1. Zheng W, Wu S, Zhang B, et al. Achieving lower radiation dose of cerebral computed tomography angiography: The role of high-concentration contrast medium. Eur J Radiol. 2026. PMID: 42150388. doi:10.1016/j.ejrad.2026.112933

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。