超快速加速合成MRI采集同时保持定量值的验证研究

Ultrafast acceleration of Synthetic MRI acquisition while maintaining quantitative values: Validation study.
PubMed 文献 · Eur J Radiol · 2026-05-19

结果显示R2、T2和PD在不同采集方法间无显著差异,且ROI内SD随采集时间缩短而降低,R2和T2在1分钟超快速采集下仍保持高度一致性。

要点

  • 1分钟超快速SyMRI采集下R2和T2值与常规方案保持高度一致。
  • R2、T2和PD在不同采集方法间总体无显著差异。
  • ROI内标准差随采集时间缩短而降低,图像噪声可能减少。

局限性

  • 仅使用3.0 T MRI和MDME序列,未涉及其他设备或序列。
  • 未在患者群体中验证,仅限健康志愿者。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: Synthetic MRI (SyMRI) is widely used in both clinical and research settings; however, its relatively long acquisition time of approximately 5-6 min has limited its applicability to acute neurological conditions in which fast imaging is critical. The aim of this study was to validate an ultrafast SyMRI acceleration algorithm and to identify quantitative parameters that maintain high agreement with conventional protocol even under such ultrafast acquisition conditions.

MATERIALS AND METHODS: Ten healthy volunteers (6 females and 4 males, median age: 27) were included. The multi-dynamic multi-echo (MDME) sequence at 3.0 T MRI were performed. Three different scan durations were used, and two types of acceleration factors-SENSE and Smart Speed AI (SSAI)-were applied, resulting in a total of three imaging protocols: 5.5-min SENSE, 4-min SSAI, and 1-min SSAI protocols. ROI measurements were performed at 15 points for each image, and quantitative values including R1, R2, T1, T2, and PD were compared. The standard deviation (SD) within the measured ROIs was also compared. Bland-Altman analysis, linear mixed-effect model, and intraclass correlation coefficient (ICC) analysis were performed.

RESULTS: Overall, no significant differences were observed between the acquisition methods for R2, T2, and PD. When evaluated separately across the 15 anatomical regions, no significant differences were observed between the acquisition methods for R2 and T2. The SD within the ROIs decreased with shorter acquisition times. Inter-reader agreement for the measurements was high.

CONCLUSION: R2 and T2 values showed high agreement even with acceleration in the ultrafast SyMRI with a 1-min acquisition time.

来源:Eur J Radiol,查看原文。版权归原作者所有。