本研究评估双对比学习GAN(DCLGAN)将短时采集重建为高质量神经PET图像,并与全时采集及2.5D U-Net比较。在FBB、FDG和FP-CIT三种示踪剂中,5分钟DCLGAN图像接近全时性能,读者评分高且SUVR/SBR一致性极佳。结果表明临床可行的扫描时间缩短是可行的,同时保持图像可解释性和定量完整性。

要点

  • 5分钟DCLGAN图像在三种示踪剂中均接近全时采集的图像质量指标。
  • 5分钟DCLGAN的SUVR/SBR与全时扫描一致性极佳(ICC≥0.995)。
  • 5分钟DCLGAN较3分钟重建图像质量更高且读者评分不降低。

局限性

  • 单中心回顾性研究。
  • 未报告诊断准确性(如敏感度、特异度)数据。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

BACKGROUND AND PURPOSE: Shortening PET/CT acquisition without degrading diagnostic or quantitative performance would improve patient comfort and scanner throughput. We evaluated a Dual-Contrastive Learning GAN (DCLGAN) for reconstructing high-quality neuro-PET images from reduced-time acquisitions across three clinically used tracers (18F-florbetaben (FBB), 18F-FDG, and 18F-FP-CIT) benchmarking against representative state-of-the-art deep learning models.

MATERIALS AND METHODS: In this single-center retrospective study of symptomatic patients undergoing neuro-PET/CT, list-mode data from full-time acquisitions (20 min for FBB; 15 min for FDG and FP-CIT) were reconstructed into 3- and 5-minute images. DCLGAN was trained to translate short-time to full-time appearance; raw images and standard 2.5D U-Net served as comparators. Image quality was quantified by normalized root-mean-square error (NRMSE), peak signal-to-noise ratio (PSNR), and structural similarity (SSIM). Two blinded nuclear medicine physicians graded clinical image quality on a five-point scale; inter-reader agreement used Cohen's kappa. Quantitative agreement of standardized uptake value ratio (SUVR) for FBB/FDG and specific binding ratio (SBR) for FP-CIT between predicted and full-time images was assessed with intraclass correlation coefficients (ICCs) and Bland-Altman analysis.

RESULTS: With DCLGAN at 5 minutes, image-quality metrics approached full-time performance across tracers: for FBB, NRMSE 0.018, PSNR 34.721 dB, SSIM 0.963; for FP-CIT, NRMSE 0.008, PSNR 41.889 dB, SSIM 0.982; and for FDG, NRMSE 0.013, PSNR 37.502 dB, SSIM 0.986. Mean reader quality scores for 5-minute DCLGAN images were high-4.48 (FBB), 4.50 (FDG), and 4.55 (FP-CIT)-with strong inter-reader agreement (κ=0.85, 0.83, 0.89, respectively). Quantitative agreement with full-time scans was excellent: ICCs for 5-minute images were 0.997 (FBB), 0.995 (FDG), and 0.996 (FP-CIT). Bland-Altman analyses showed minimal bias (mean bias -0.004 for FBB, 0.012 for FDG, -0.029 for FP-CIT) with narrow limits of agreement. Compared with 3-minute reconstructions, 5-minute DCLGAN images demonstrated significantly higher PSNR/SSIM and lower NRMSE (p<0.001), while maintaining or improving reader scores and quantitative agreement.

CONCLUSIONS: Across FBB, FDG, and FP-CIT, DCLGAN enabled 5-minute neuro-PET acquisitions that appeared to maintain reader-rated quality and showed high agreement in SUVR/SBR values with full-time scans. These findings suggest that clinically practical scan-time reduction may be feasible while maintaining interpretability and quantitative integrity in this cohort.