动态增强MRI与MAG-3肾闪烁显像评估儿童分肾功能对比

Comparison of dynamic contrast-enhanced MRI versus MAG-3 scintigraphy for differential renal function assessment in pediatric patients.
PubMed 文献 · Pediatr Radiol · 2026-04-29

回顾性纳入23例患儿,比较动态增强MRI与MAG-3肾闪烁显像测得的分肾功能(DRF)。两种方法高度线性相关、一致性良好且无系统性偏倚,但测量变异度提示在手术决策阈值附近需谨慎解读。建议肾积水肾脏的MAG-3分析采用C形本底校正以优化方法间一致性。

要点

  • DCE-MRI与MAG-3测得的DRF高度相关(r=0.872),一致性相关系数0.870
  • 78.3%(18/23)病例的DRF差异在预设临床可接受范围±10%内
  • 手术决策阈值附近的DRF值需谨慎解读,因测量变异度较大

核心结论

  • DCE-MRI与MAG-3闪烁显像在儿童分肾功能评估中呈强线性相关和良好一致性,未检测到有临床意义的系统性或比例偏倚。
  • 观察到的测量变异度提示:当DRF值接近手术决策阈值时,比较两种方法的结果需谨慎。
  • 对于肾积水肾脏,推荐MAG-3分析采用C形本底校正,以优化方法间的一致性。

研究设计与样本

  • 回顾性研究,纳入23例患儿(年龄3个月至20岁,平均8.3±7.4岁),在6个月内完成MR尿路造影协议中的DCE-MRI和MAG-3检查。
  • DCE-MRI采用黄金角径向动态堆叠星采集与GRASP重建,并用双室示踪动力学模型分析肾滤过率。
  • 以MAG-3闪烁显像为参考,分别采用默认本底(n=23)和C形本底(n=16,有原始数据者)分析DRF,用Bland-Altman分析、Pearson相关和一致性相关系数评估一致性。

诊断标准与关键数值

  • 默认MAG-3本底分析(n=23):平均DRF差异(MAG-3减DCE-MRI)为0.4%(95%CI -3.3%至4.2%),95%一致性界限为-16.7%至+17.5%,一致性相关系数0.870(95%CI 0.718-0.942)。
  • 预设临床可接受范围±10%内,78.3%(18/23)病例的DRF差异落在此范围内,平均绝对差异为6.1%。
  • C形本底校正亚组(n=16):平均差异-2.3%(95%CI -6.7%至2.0%),一致性界限-18.3%至+13.6%,一致性相关系数0.845(95%CI 0.614-0.943)。
  • Pearson相关在两种分析中均强:n=23时r=0.872(95%CI 0.719-0.945),n=16时r=0.866(95%CI 0.648-0.953),均P<0.001。

临床应用与阅片要点

  • DCE-MRI可作为无辐射、高分辨率的替代方法评估儿童分肾功能,尤其适用于需多次随访的患儿。
  • 阅片报告DRF时,应意识到DCE-MRI与MAG-3的测量差异可能达±10%以上,在手术决策阈值附近需结合临床和其他影像综合判断。
  • 对肾积水肾脏,建议MAG-3分析采用C形本底校正,以提高与DCE-MRI的一致性。

局限性

  • 回顾性研究,样本量较小(n=23),可能存在选择偏倚。
  • C形本底校正亚组仅16例有原始数据,亚组分析效能有限。
  • 未报告重复测量的一致性,测量变异度可能受操作者影响。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

BACKGROUND: Measuring differential renal function (DRF) is critical for managing children with congenital and acquired urological diseases. MAG-3 renal scintigraphy is the most widely used clinical technique; dynamic contrast-enhanced MRI (DCE-MRI) provides a radiation-free and high-resolution alternative method for assessing DRF.

OBJECTIVE: To evaluate the agreement between DRF values derived from DCE-MRI and MAG-3 scintigraphy-derived DRF in pediatric patients.

MATERIALS AND METHODS: This retrospective study included 23 pediatric patients (age range, 3 months-20 years; mean±SD, 8.3±7.4 years) who underwent DCE-MRI embedded in MR urography (MRU) protocol and MAG-3 studies within 6 months. DCE-MRI was performed using a golden-angle radial dynamic stack-of-stars acquisition with golden-angle sparse parallel (GRASP) reconstruction, and analyzed the renal filtration rate with a two-compartment tracer kinetic model. DRF values from DCE-MRI were compared with those from MAG-3, using both default (n=23) and C-shaped background regions for MAG-3 analysis, where raw data were available (n=16). Agreement was evaluated using Bland-Altman analysis, Pearson and concordance correlations, and paired statistical tests.

RESULTS: With default MAG-3 background analysis (n=23), mean DRF difference (MAG-3 minus DCE-MRI) was 0.4% (95% CI, -3.3% to 4.2%), with 95% limits of agreement of -16.7% to +17.5%. Concordance correlation coefficient (CCC) was 0.870 (95% CI, 0.718-0.942). Within the pre-defined clinically acceptable margin of ±10%, 78.3% (18/23) of cases fell, with a mean absolute difference of 6.1%. With C-shaped background subtraction in the subset with available raw MAG-3 data (n=16), mean difference was -2.3% (95% CI, -6.7% to 2.0%), limits of agreement -18.3% to +13.6%, and CCC 0.845 (95% CI, 0.614-0.943). Pearson correlation was strong in both analyses (n=23: r=0.872, 95% CI, 0.719-0.945; n=16: r=0.866, 95% CI, 0.648-0.953; both P<0.001). No clinically significant systematic or proportional bias was detected in either analysis.

CONCLUSION: DCE-MRI demonstrates strong linear association and concordance with MAG-3 scintigraphy for pediatric differential renal function assessment with no clinically significant systematic bias, though the observed measurement variability indicates caution is warranted when comparing values near surgical decision thresholds. C-shaped background correction is recommended for MAG-3 in hydronephrotic kidneys to optimize inter-modality agreement.

来源:Pediatr Radiol,查看原文。版权归原作者所有。