反式全肩关节置换术后X线片与CT的DSA和LSA角度测量比较

Comparison of DSA and LSA angle measurements in postoperative radiographs and computed tomography of patients with reverse total shoulder arthroplasty.
PubMed 文献 · Skeletal Radiol · 2026-04-23

本研究比较了80例患者88个肩在反式全肩关节置换术后X线片与CT多平面重建上测量的肩峰远端化角(DSA)和肩峰外侧化角(LSA)。结果显示两种方法相关性很好,LSA平均差异2°,DSA平均差异3.5°,两种角度在X线片和CT上均有较强的观察者间信度。

要点

  • X线片与CT测量的LSA和DSA高度相关,临床可互参考。
  • LSA的观察者间信度X线片ICC=0.691,CT ICC=0.721。
  • DSA的观察者间信度X线片ICC=0.858,CT ICC=0.768。

诊断标准与关键数值

  • LSA的X线片与CT测量平均差异为2°(SD±3.6°,范围-8.5至5°,p<0.001),Spearman's Rho=0.876,ICC在X线片为0.691、CT为0.721。
  • DSA的X线片与CT测量平均差异为3.5°(SD±4.8°,范围-9.5至12°,p<0.001),Spearman's Rho=0.855,ICC在X线片为0.858、CT为0.768。

临床应用与阅片要点

  • 日常测量DSA和LSA时,X线片与CT结果可相互参考,但需注意DSA的平均差异略大(3.5°)。
  • 观察者间信度DSA在X线片上较高(ICC=0.858),LSA在两种方法上均为中等至良好(ICC 0.691-0.721)。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: The distalization shoulder angle (DSA) and lateralization shoulder angle (LSA) are radiographic measurements used to assess the position of the implant and its impact on shoulder function in reverse shoulder arthroplasty (RSA). The present study aims to determine (a) whether LSA and DSA measurements obtained from radiographs and CT scans of the same patients yield comparable results and (b) the inter-observer reliability of radiographic and CT measurements.

MATERIALS AND METHODS: LSA and DSA measurements of 88 shoulders in 80 patients were performed on radiographs and multiplanar reconstructions of CT scans. A paired t-test was used to compare angle measurements both between imaging modalities (radiograph versus CT) and between observers. Inter-method correlation was evaluated using Spearman's Rho and inter-observer agreement was assessed using the intraclass correlation coefficient (ICC).

RESULTS: There was a very good inter-method correlation for LSA (Spearman's Rho = 0.876), with the total mean differences between radiographic and CT measurements being 2° (SD ± 3.6°, range -8.5 to 5°, p < 0.001). The ICC for LSA was 0.691 for radiographs and 0.721 for CT scans. There was also a very good inter-method correlation for DSA (Spearman's Rho = 0.855), with the total mean difference between radiographic and CT measurements being 3.5° (SD ± 4.8°, range -9.5 to 12°, p < 0.001). The ICC for DSA was 0.858 for radiographs and 0.768 for CT scans.

CONCLUSION: CT-derived measurements of the DSA and LSA demonstrate high reproducibility, showing strong agreement with radiographic measurements. Strong inter-observer reliability can be expected for both angles in both radiographic and CT assessments.

来源:Skeletal Radiol,查看原文。版权归原作者所有。

引用

  1. Mitrousias V, Brandariz R, Dor J, et al. Comparison of DSA and LSA angle measurements in postoperative radiographs and computed tomography of patients with reverse total shoulder arthroplasty. Skeletal Radiol. 2026. PMID: 42024153. doi:10.1007/s00256-026-05226-y

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。