EOS双平面成像评估肩胛胸壁运动障碍:一种新的简化方法及可行性研究

Evaluation of scapulothoracic dyskinesis (SD) using EOS biplanar imaging: A new simplified method and feasibility study.
PubMed 文献 · Skeletal Radiol · 2026-04-23

本研究利用肩胛胸壁骨模型模拟正常及三种Kibler分型的肩胛胸壁运动障碍,采用低剂量EOS双平面成像在七个不同体位下进行三维测量。结果表明所有解剖标志识别与测量均可行,观察者间和观察者内可重复性除内旋(0.75)外均大于0.80,提示该方法可能成为三维评估肩胛骨位置的有前景工具,但需临床验证。

要点

  • EOS双平面成像可在骨模型上可重复地测量肩胛胸壁位置。
  • 除内旋外,各参数观察者间和观察者内可重复性均大于0.80。
  • 该方法尚需临床环境验证,暂不能直接用于患者诊断。

诊断标准与关键数值

  • 观察者间和观察者内可重复性在所有参数中均大于0.80,但内旋参数为0.75。
  • 观察者间角度测量的平均差异为0.3±2.8°,距离测量的平均差异为0.1±0.5 cm。

临床应用与阅片要点

  • 在骨模型上,EOS双平面成像可用于识别七个解剖标志并进行三维肩胛胸壁测量。
  • 内旋参数的可重复性相对较低(0.75),临床解读时需谨慎。

局限性

  • 研究仅在骨模型上进行,尚未在临床患者中验证。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

BACKGROUND: Scapulothoracic dyskinesis (SD) is a common functional disorder of the shoulder characterized by abnormal positioning or mobility of the scapula. Kibler's classification describes three catagories of SD. Clinical and radiological diagnosis remain challenging, and no simple 3D assessment method is currently available.

PURPOSE: The objectives of this study were (1) to determine whether scapular positioning on bone models simulating SD conditions can be assessed using low-dose EOS biplanar imaging, and (2) to evaluate the inter- and intra-observer reproducibility of this method.

METHODS: A full-scale scapulothoracic bone model including bilateral scapulae, thorax, and spine was used. Scapular positions were adjusted to simulate one normal condition and three SD patterns corresponding to Kibler's classification. EOS imaging was performed in seven different positions for each model. Seven anatomical landmarks were identified on stereoradiographs to obtain 3D scapulothoracic measurements. Twenty-eight sets of EOS measurements were analyzed by two independent observers. Inter- and intra-observer reproducibility were assessed.

RESULTS: Landmark identification and scapulothoracic measurements were feasible in all cases. Inter- and intra-observer reproducibility were > 0.80 for all parameters except internal rotation (0.75). The mean difference between observers was 0.3 ± 2.8° for angular measurements and 0.1 ± 0.5 cm for distance measurements.

CONCLUSION: EOS biplanar imaging enables reproducible measurement of scapulothoracic positioning on dry bone using a simplified method. These findings suggest that this approach may be a promising tool for 3D assessment of scapular positioning, although validation in clinical settings is required.

来源:Skeletal Radiol,查看原文。版权归原作者所有。