超声与机械装置结合IMU评估儿童胫骨及胫腓骨扭转指数的可靠性

Reliability of Tibial and Tibiofibular Torsion Indices Using Ultrasound and Mechanical Devices With an Inertial Measurement Unit (IMU) in Children.
PubMed 文献 · Ultrasound Med Biol · 2026-04-18

前瞻性研究纳入20名7-9岁健康儿童(40根胫骨),评估超声胫骨扭转指数(TTI)和机械胫腓骨扭转指数(TFTI)的观察者内与观察者间信度。TTI观察者内ICC为0.98、观察者间为0.86;TFTI分别为0.85和0.70;两种方法相关性弱(r=0.35),平均差异14.7°。两者可重复性均良好,超声TTI观察者间信度更优且儿童接受度高。

要点

  • 超声TTI观察者间信度良好(ICC=0.86),优于机械TFTI(ICC=0.70)。
  • TTI与TFTI相关性弱(r=0.35),不可直接互换,平均差异14.7°。
  • 测量误差SEM为1.6°-3.8°,最小可检测变化SDC95为4.4°-10.4°。

核心结论

  • 超声TTI和机械TFTI在儿童中均具有可重复性,但超声TTI观察者间信度更优,可能成为儿童友好的无辐射监测胫骨扭转的选择。

研究设计与样本

  • 前瞻性信度研究,纳入20名7-9岁健康儿童共40根胫骨,由两名独立观察者分不同日重复测量。
  • TTI采用超声结合探头上的紧凑型惯性测量单元(IMU),TFTI采用两个机械装置序贯使用单个IMU。
  • 信度指标包括ICC(2,1)及95%CI、测量标准误(SEM)和最小可检测变化(SDC95),并评估儿童接受度。

诊断标准与关键数值

  • TTI观察者内ICC(2,1)=0.98,观察者间ICC(2,1)=0.86;TFTI观察者内ICC(2,1)=0.85,观察者间ICC(2,1)=0.70。
  • SEM范围为1.6°-3.8°,SDC95范围为4.4°-10.4°。
  • TTI与TFTI相关性弱(r=0.35;ρ=0.30),Bland-Altman平均差异14.7°并存在比例偏倚。
  • 儿童接受度高:舒适度中位数5、无痛5、愿意重复4/5(Likert量表)。

临床应用与阅片要点

  • 超声TTI可作为儿童胫骨扭转的无辐射监测工具,但需注意观察者间信度仍非完美(ICC=0.86)。
  • TTI与TFTI相关性弱且平均差异达14.7°,临床解读时不可将两种方法结果直接互换或等同。
  • SEM和SDC95提示测量误差范围,判读变化时应考虑最小可检测变化(4.4°-10.4°)以避免过度解读。
  • 儿童接受度高,适合重复随访,但需标准化 landmarks 和 IMU 操作以减小观察者间变异。

局限性

  • 研究样本量小(20名儿童,40根胫骨),仅纳入7-9岁健康儿童,外推至其他年龄或患者群体需谨慎。
  • TTI与TFTI相关性弱,但未与金标准(如CT)比较,两者准确性未知。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: To assess the intra- and inter-observer reliability of two non-invasive tibial torsion indices in children-the ultrasound-based Tibial Torsion Index (TTI) and the mechanically based Tibiofibular Torsion Index (TFTI)-and to examine their agreement and child acceptability.

MATERIALS AND METHODS: Prospective reliability study in 20 healthy children aged 7-9 years (40 tibiae). TTI was obtained with ultrasound using standardized landmarks and a compact inertial measurement unit (IMU) mounted on the probe; TFTI with two mechanical devices using a single IMU sequentially. Two independent observers performed assessments (Observer 1 on Day 1 and Day 2; Observer 2 on Day 2). Reliability metrics included intraclass correlation coefficients (ICC(2,1), 95% CI), standard error of measurement (SEM) and smallest detectable change (SDC95). Intrasession variability was expressed as within-session SD. Agreement was evaluated using correlation, Bland-Altman plots and regression analyses. Child acceptability regarding pain, comfort and willingness to repeat was recorded on a Likert-type scale supported by a faces-based visual aid.

RESULTS: TTI showed excellent intra-observer (ICC(2,1) = 0.98) and good inter-observer reliability (ICC(2,1) = 0.86). TFTI showed good intra-observer (ICC(2,1) = 0.85) and moderate inter-observer reliability (ICC(2,1) = 0.70). SEM ranged 1.6°-3.8°; SDC95 4.4°-10.4°. Correlation between TTI and TFTI was weak (r = 0.35; ρ = 0.30); Bland-Altman indicated a mean difference of 14.7° with proportional bias. Acceptability was high (median comfort 5, pain-free 5, willingness-to-repeat 4/5).

CONCLUSION: Both indices were reproducible. The ultrasound-based TTI demonstrated superior inter-observer reliability and may represent a child-friendly, radiation-free option for monitoring tibial torsion in pediatric practice.