体模校准对放射密度测量准确性的影响:骨肉瘤化疗反应评估的启示

Phantom Calibration's Impact on Measuring Radiodensity Changes: Implications for Evaluating Osteosarcoma Response.
PubMed 文献 · Acad Radiol · 2026-04-10

研究采用五个新鲜冷冻尸体下肢,在股骨远端置入已知放射密度的棒状物模拟肿瘤,比较有无铝阶梯体模校准对放射密度减影准确性的影响。

要点

  • 在相同成像设置下,不使用体模校准的放射密度测量准确性更高。
  • 侧位膝关节X线片及特定成像设置(4 mAs-70 kVp、8 mAs-70 kVp、4 mAs-80 kV)可提高准确性。
  • 骨密度/BMI变化可能影响测量准确性,但具体关系需进一步研究。

局限性

  • 研究为尸体离体研究,不能完全模拟活体生理和病理环境。
  • 样本量较小(5个尸体下肢),可能影响结果的普遍性。
  • 未评估不同成像技术(如变化kVp/mAs)下校准的效果。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

RATIONALE AND OBJECTIVES: Improve radiodensity measurements through a controlled, cadaveric study with the primary hypothesis that using high-resolution digital radiographs and image calibration with an aluminum step phantom would be more accurate than in the absence of a phantom. Ultimately, the goal is to improve evaluation of osteosarcoma response to chemotherapy prior to resection using radiographs.

MATERIALS AND METHODS: Five fresh-frozen cadaveric lower extremities were imaged with computed tomography (CT) and radiography (anteroposterior (AP) and lateral) with varying mAs/kVp before and after distal femoral instrumentation with 4 cm2 rods of known radiodensities, simulating tumors of varying density, and a commercial radiographic calibration step phantom. Using Materials Laboratory for Digital Imaging and Communications in Medicine (DICOM) image analysis, grayscale measurements were obtained for: rods in/out of bone, normal bone proximal to instrumentation, air, and all 16 calibration phantom steps. Radiodensity subtraction accuracy was compared for calibration with all phantom steps, +/- bone or air subtraction. Accuracy was compared using Student T-test and one-way analysis of variance with Tukey HSD post-hoc test.

RESULTS: Mean accuracy for Cohort 1 (no calibration) (.12, standard deviation [SD] 0.11, N = 120) was better than that of Cohorts 2-4 (calibration) (0.39, SD 0.54-0.60, N = 1920) (p < 0.001, t = 5.4689). Lateral knee radiographs, specific imager settings (4 mAs-70 kVP, 8 mAs-70 kVp, and 4 mAs-80 kV), and variations in bone density/BMI improved accuracy (p < .001).

CONCLUSION: When measuring radiodensity changes using high-quality, digital images with identical settings, image calibration worsened accuracy, likely by introducing an additional source of potential error. Further work is required to determine if calibration will improve accuracy for images obtained with varying technique.

来源:Acad Radiol,查看原文。版权归原作者所有。

引用

  1. McNamara MA, Mazza PM, Hebert KA, et al. Phantom Calibration's Impact on Measuring Radiodensity Changes: Implications for Evaluating Osteosarcoma Response. Acad Radiol. 2026. PMID: 41966847. doi:10.1016/j.acra.2026.03.045