3-T与5-T钆塞酸二钠增强MRI评估HCC的个体内对比初步研究
该研究前瞻性纳入41例疑似HCC患者,个体内对比3-T与5-T Gd-EOB-DTPA增强MRI。5-T在增强期主观评分更高,DWI病灶边缘清晰度更优,增强包膜检出率更高。5-T结合ACS技术可提升图像质量和影像特征评估。
要点
- 5-T增强期图像主观评分优于3-T,但伪影无差异。
- 5-T DWI病灶边缘清晰度更高,对比度更优。
- 5-T对HCC增强包膜的检出率高于3-T。
核心结论
- 5-T Gd-EOB-DTPA增强MRI在增强期图像质量和增强包膜检出方面优于3-T。
- 5-T结合ACS技术有望进一步提升图像质量和影像特征评估。
研究设计与样本
- 前瞻性个体内对比研究,纳入41例疑似HCC患者。
- 患者均接受5-T和3-T Gd-EOB-DTPA增强MRI。
- 5-T肝胆期成像同时使用ACS和PI技术。
- 由两名放射科医师进行图像质量与影像特征的主观和定量评估。
关键影像表现/征象
- 5-T增强期主观评分高于3-T(均p<0.05),但图像伪影无差异。
- 5-T DWI病灶边缘清晰度主观评分高于3-T(p=0.021)。
- 5-T DWI对比度高于3-T(p<0.05)。
- 5-T对增强包膜的检出率高于3-T(p=0.016)。
- 5-T使用ACS时肝胆期瘤周低信号检出率更高(p=0.015)。
临床应用与阅片要点
- 5-T MRI可提高HCC增强包膜的检出,有助于诊断。
- 肝胆期使用ACS技术可改善图像质量和瘤周低信号评估。
局限性
- 样本量较小(41例),为初步研究。
- 单中心前瞻性研究,结果需外部验证。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND: We aimed to evaluate the utility of 5-T Gadolinium-ethoxybenzyl diethylenetriaminepentaacetic acid (Gd-EOB-DTPA)-enhanced magnetic resonance imaging (MRI) by intraindividual comparison with 3-T MRI, with a focus on the image quality and diagnosis of hepatocellular carcinoma (HCC).
METHODS: We prospectively enrolled 41 patients with suspected HCC who underwent dynamic Gd-EOB-DTPA-enhanced MRI using both 5-T and 3-T scanners. Artificial intelligence-assisted compressed sensing (ACS) and parallel imaging (PI) were both used for hepatobiliary phase (HBP) imaging at 5-T. Two radiologists performed qualitative and quantitative assessments of image quality and evaluations of imaging features. Wilcoxon signed-rank, paired χ2, and Cochran Q tests as well as intraclass correlation coefficients and the Cohen κ value were used.
RESULTS: All subjective image quality scores ranged from good to excellent. The subjective scores of contrast-enhanced phases for 5-T images were higher than those for 3-T images (all p < 0.05), except for image artifacts. For diffusion-weighted imaging (DWI), the subjective scores of the clarity of the lesion margins on 5-T images were higher than those on 3-T images (p = 0.021). Quantitative measures were also greater for 5-T images (all p < 0.05). For DWI, the contrast ratio (CR) at 5-T was greater (p < 0.05). Subjective and quantitative assessments of HBP imaging were higher with ACS (all p < 0.05). The detection rate of enhancing capsule was greater for 5-T images (p = 0.016), as was the rate of peritumoral hypointensity on the HBP images at 5-T using ACS (p = 0.015).
CONCLUSIONS: Compared with 3-T MRI, liver dynamic Gd-EOB-DTPA-enhanced 5-T MRI demonstrated superior image quality for contrast-enhanced phases and greater sensitivity in detecting enhancing capsule in HCC. The integration of 5-T MRI and ACS technology has the potential to further improve image quality and the assessment of imaging features.
RELEVANCE STATEMENT: Gd-EOB-DTPA-enhanced 5-T MRI provides promising potential for accurate HCC evaluation.