新型肩挂式辐射防护屏在介入手术中的眼辐射防护效率与工效学性能

Ocular radiation protection efficiency and ergonomic performance of a novel Shoulder-Mounted radiation shield for interventional procedures.
PubMed 文献 · Eur J Radiol · 2026-03-30

本研究通过体模实验比较新型肩挂式辐射防护屏与传统头挂式防护屏在冠状动脉造影中的眼辐射剂量与工效学性能;肩挂式联合辅助屏蔽平均降低眼剂量94.2%,优于头挂式的90.4%(p<0.01),且降低颈椎椎间盘应力,佩戴舒适度更佳而操作灵活性相当。

要点

  • 肩挂式防护屏联合辅助屏蔽平均降低眼剂量94.2%,优于头挂式联合辅助屏蔽的90.4%。
  • 肩挂式较头挂式降低头部负荷及颈椎椎间盘应力,水平视图最大降51.0%,45°下视最大降24.1%。
  • 肩挂式佩戴舒适度优于头挂式(p<0.001),操作灵活性相当(p=0.301)。

诊断标准与关键数值

  • 肩挂式防护屏联合辅助屏蔽较无屏蔽平均降低眼剂量94.2%(p<0.01)。
  • 头挂式防护屏联合辅助屏蔽较无屏蔽平均降低眼剂量90.4%(p<0.01)。
  • 肩挂式较头挂式降低颈椎椎间盘应力水平视图最大51.0%,45°下视最大24.1%。
  • 用户调查肩挂式佩戴舒适度优于头挂式(p<0.001),操作灵活性相当(p=0.301)。

局限性

  • 本研究为体模研究,非临床实际操作者测量。
  • 用户调查样本量及人群特征未报告。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: The importance of ocular radiation protection for interventionalists is increasingly recognized. This study compares ocular radiation protection efficiency and ergonomic performance of a novel shoulder-mounted radiation shield with those of a conventional head-mounted shield, aiming to provide a new and optimized ocular radiation protection solution for interventional procedures.

METHODS: In this phantom study, the simulated operator's ocular doses were measured under various radiation protection configurations during coronary angiography. Head load was measured using a pressure transducer, and finite element analysis simulated cervical intervertebral disc stress. A user survey assessed wearing comfort and operational flexibility of the two shields.

RESULTS: Compared to no shielding, the novel shoulder-mounted shield combined with the ancillary shields achieved an average ocular dose reduction of 94.2%, surpassing the average reduction of 90.4% attained by the conventional head-mounted shield combined with the ancillary shields (both p < 0.01). Compared to the head-mounted shield, the shoulder-mounted shield effectively reduced the load on the head and cervical intervertebral discs, decreasing the stresses in the cervical intervertebral discs by up to 51.0% in the horizontal view and by up to 24.1% in the 45° downward view. The user survey indicated the shoulder-mounted shield provided better wearing comfort than the head-mounted shield (p < 0.001) with comparable operational flexibility (p = 0.301).

CONCLUSIONS: The novel shoulder-mounted shield could provide better ocular radiation protection efficiency and ergonomic performance compared to the conventional head-mounted shield. This study presents a new and optimized ocular radiation protection solution for interventional procedures that effectively balances radiation safety and wearing comfort.

来源:Eur J Radiol,查看原文。版权归原作者所有。

引用

  1. Qian H, Gao Y, Xu Y, et al. Ocular radiation protection efficiency and ergonomic performance of a novel Shoulder-Mounted radiation shield for interventional procedures. Eur J Radiol. 2026. PMID: 41946019. doi:10.1016/j.ejrad.2026.112832

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。