三维多回波稳态采集与STIR序列在臂丛MR神经成像中的比较
50例受试者在3T下行单侧臂丛MR神经成像,先后采集非增强STIR-FSE与MENSA-NERVE序列并回顾性分析。MENSA-NERVE血管抑制优于STIR-FSE(P < .001),脂肪抑制STIR-FSE更优(P = .05至P < .001),MENSA-NERVE对肩胛上神经和腋神经显示更优(P = .004至P < .001)。
要点
- MENSA-NERVE血管抑制优于非增强STIR-FSE(P < .001)。
- STIR-FSE脂肪抑制优于MENSA-NERVE,MENSA-NERVE脂肪抑制不全主要见于肺尖。
- MENSA-NERVE对肩胛上神经和腋神经显示优于STIR-FSE(P = .004至P < .001)。
核心结论
- MENSA-NERVE较非增强STIR-FSE血管抑制和神经显示更优,但在颈肩交界和肺尖脂肪抑制不够稳健。
研究设计与样本
- 回顾性分析50例受试者(24例女性,中位年龄34岁,范围14-73岁)在3T下行单侧臂丛MR神经成像,先后采集非增强STIR-FSE与MENSA-NERVE序列,由2名放射科医师评估。
关键影像表现/征象
- MENSA-NERVE脂肪抑制不全遮蔽神经显示,主要仅见于肺尖。
- MENSA-NERVE在颈肩交界和肺尖脂肪抑制不够稳健。
诊断标准与关键数值
- MENSA-NERVE血管抑制优于STIR-FSE(P < .001)。
- 脂肪抑制评估STIR-FSE优于MENSA-NERVE(P = .05至P < .001)。
- MENSA-NERVE对肩胛上神经和腋神经显示优于STIR-FSE(P = .004至P < .001)。
- 两序列评估者间对可见病变的 agreement 为0.74-0.92(好至优)。
- 序列间 agreement 为0.80-0.88(优)。
临床应用与阅片要点
- 作者结论认为MENSA-NERVE血管抑制和神经显示优于非增强STIR-FSE,但颈肩交界和肺尖脂肪抑制不够稳健。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND AND PURPOSE: The STIR-FSE pulse sequence is commonly used for 3D MR neurography (MRN) of the brachial plexus (BP) due to its low B0 susceptibility and robust fat suppression. An alternative 3D acquisition is a multiecho in steady-state acquisition (MENSA-NERVE) sequence that typically provides more robust vascular suppression but poorer fat suppression due to its use of a water excitation pulse.
MATERIALS AND METHODS: A total of 50 subjects (24 women, median age [range]: 34 years [14-73 years]) underwent unilateral BP MRN at 3T where noncontrast STIR-FSE and MENSA-NERVE were sequentially acquired and retrospectively analyzed. Sequences were assessed by 2 radiologists for vascular suppression, fat suppression, and visualization of the BP nerve roots, trunks, and cords, as well as of the suprascapular nerve and axillary nerve. Pooled measurements were analyzed using the Wilcoxon signed-rank test, and agreement proportion was used to assess interrater/intersequence diagnostic agreement.
RESULTS: MENSA-NERVE outperformed STIR-FSE for vascular suppression (P < .001). Fat suppression assessments favored STIR-FSE over MENSA-NERVE (P = .05 to P < .001), with incomplete fat suppression obscuring nerve visualization in MENSA-NERVE, primarily observed only at the lung apex. MENSA-NERVE provided superior visualization of the suprascapular and axillary nerves (P = .004 to P < .001). Interrater agreement of visualized pathology was good to excellent (agreement = 0.74-0.92) for both sequences. Intersequence agreement was also excellent (agreement = 0.80-0.88).
CONCLUSIONS: MENSA-NERVE provided superior vascular suppression and nerve conspicuity compared with noncontrast STIR-FSE and with less robust fat suppression at the neck-shoulder junction and lung apex.