使用多透镜探头进行人颈动脉4D宽场超快超声成像

4D Wide-Field Ultrafast Ultrasound of the Human Carotid Using a Multi-Lens Probe.
PubMed 文献 · Ultrasound Med Biol · 2026-03-24

本研究在体外流动体模和人颈动脉中验证1MHz大孔径多透镜探头的4D超快多普勒成像性能,结果显示测得的管径与商用超声设备一致,可检测相反流向并捕捉颈动脉搏动频谱,体外和体内视野分别达41×69×41 mm³和39×103×32 mm³,穿透深度分别达8 cm和6 cm。

要点

  • 大孔径多透镜探头可实现颈动脉4D宽场超快多普勒成像,体内视野39×103×32 mm³、穿透深度6 cm。
  • 体模中管径测量结果与商用超声设备一致,提示可获取结构信息。
  • 系统对血流方向敏感,可检测体模内相反流向并捕捉颈动脉搏动频谱。

核心结论

  • 大孔径多透镜探头可实现人颈动脉宽场4D超快多普勒成像,作者认为其在临床环境中有很大潜力改善血管成像。

研究设计与样本

  • 体外流动体模(两支管,水-纤维素混合物灌注)与人颈动脉在体研究,使用1MHz大孔径多透镜探头采集射频数据,采用3D延迟叠加波束合成(10至12个发散波相干复合、透镜校正、动态聚焦)及奇异值分解滤波获取能量多普勒、彩色与频谱多普勒数据;体模管径与常规线阵探头参考测量比较。

诊断标准与关键数值

  • 能量多普勒容积测得的管径与优质商用超声设备(Aixplorer,Supersonic Imagine)测量结果一致。
  • 体外视野41×69×41 mm³,穿透深度达8 cm;体内视野39×103×32 mm³,穿透深度达6 cm。

临床应用与阅片要点

  • 作者认为该大孔径多透镜探头在临床环境中有很大潜力改善血管成像。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: To extend the use of a large multi-lens probe technology to 4D ultrafast Doppler imaging and validate its performance both in vitro using a flow phantom and in vivo on the human carotid artery.

METHODS: Radio-frequency data were acquired using a 1MHz large-aperture multi-lens probe on a tilted flow phantom with two tubes perfused by a water-cellulose mixture, and on a human carotid artery in vivo with contrast agents. 3D delay-and-sum beamforming with coherent compounding of 10 to 12 diverging waves, lens correction, and dynamic focusing was performed, followed by singular value decomposition (SVD) filtering to obtain Power Doppler data, Color and spectral Doppler data. Tube diameters were measured from Power Doppler volumes and compared with reference measurements taken with a conventional linear probe.

RESULTS: Tube diameters measured with the probe were consistent with those using a premium image quality commercial echographic device (Aixplorer, Supersonic Imagine), demonstrating the capability to retrieve structural information. The multi-lens system showed sensitivity to flow orientation and successfully detected opposite flow directions within the phantom. Doppler spectra captured carotid pulsatility, illustrating the ability to quantify physiologically relevant flow dynamics. The large-aperture design achieved an extended field of view of 41 × 69 × 41 mm³ in the phantom and 39 × 103 × 32 mm³ in vivo, with penetration depths of up to 8 cm and 6 cm, respectively.

CONCLUSION: The large multi-lens probe enables a wide-field of view 4D ultrafast Doppler imaging of the carotid in humans and holds strong potential to improve vascular mapping in a clinical setting.

来源:Ultrasound Med Biol,查看原文。版权归原作者所有。