文献速递乳腺CT

乳腺癌保乳术后放疗中光学表面引导患者摆位准确性的评估

Evaluation of patient setup accuracy using optical surface guidance in postoperative radiotherapy for patients with breast cancer after breast-conserving surgery.
PubMed 文献 · Br J Radiol · 2026-01-01

前瞻性队列研究纳入125例保乳术后放疗患者,按手术侧别及是否使用热塑固定膜分组,每分次行CBCT验证摆位误差。结果显示无论左右侧,不使用固定膜均可改善摆位重复性;右侧保乳病例所需PTV外扩边界大于左侧。

要点

  • 保乳术后SGRT摆位中,省略热塑固定膜可改善摆位重复性。
  • 右侧保乳病例所需PTV外扩边界大于左侧保乳病例。
  • 固定膜使用与否及手术侧别均影响分次间摆位误差。

核心结论

  • 在保乳术后SGRT流程中,省略热塑固定膜可改善摆位重复性;右侧保乳病例所需PTV外扩边界大于左侧病例。

研究设计与样本

  • 前瞻性队列研究,纳入2021年1月至2025年6月在本中心接受保乳术后放疗的125例患者,按手术侧别(左/右)和固定方式(使用/不使用热塑固定膜)分组;SGRT摆位后每分次行CBCT,共获868次CBCT扫描,分析系统误差和随机误差并计算三维PTV外扩边界。

诊断标准与关键数值

  • 使用固定膜时,左侧病例X方向摆位误差0.04±0.25 cm,右侧为-0.05±0.30 cm(P<.001)。
  • 使用固定膜时,左侧病例Y方向摆位误差-0.09±0.26 cm,右侧为0.02±0.30 cm(P<.001)。
  • 不使用固定膜时,左侧病例X方向摆位误差-0.05±0.22 cm,右侧为0.01±0.27 cm(P=.017)。
  • 不使用固定膜时,左侧病例Y方向摆位误差0.03±0.23 cm,右侧为0.08±0.27 cm(P=.039)。
  • 无论侧别,固定膜使用与否对三个方向误差均有显著影响(均P<.05);所需PTV外扩边界(cm):左侧保乳使用固定膜(0.51, 0.51, 0.54),不使用固定膜(0.47, 0.49, 0.57);右侧保乳使用固定膜(0.64, 0.63, 0.55),不使用固定膜(0.59, 0.62, 0.53)。

临床应用与阅片要点

  • 本研究表明保乳术后SGRT摆位中省略热塑固定膜可改善摆位重复性。
  • 右侧保乳病例所需PTV外扩边界大于左侧,制定计划时需考虑侧别差异。
  • 研究为保乳术后全乳放疗的患者体位与固定技术提供了临床建议。

局限性

  • 原文未报告局限性

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: To evaluate the clinical application of AlignRT surface-guided radiotherapy (SGRT) by quantifying interfractional setup errors in patients with breast-conserving surgery (BCS) stratified by the surgical side and use of thermoplastic immobilization masks.

METHODS: This prospective cohort study enrolled 125 patients with BCS treated at our center between January 2021 and June 2025. Patients were grouped by the surgical side (left/right) and immobilization method (with/without thermoplastic mask). After SGRT setup, cone-beam computed tomography (CBCT) was acquired for each fraction. Systematic and random errors were analyzed and 3-dimensional planning target volume (PTV) margins were calculated.

RESULTS: In total, 868 CBCT scans were obtained. With immobilization masks, left-sided patients exhibited significantly different setup errors from right-sided patients in the X (0.04 ± 0.25 cm vs. -0.05 ± 0.30 cm, P < .001) and Y (-0.09 ± 0.26 cm vs.0.02 ± 0.30 cm, P < .001) directions. Without immobilization masks, differences persisted in X (-0.05 ± 0.22 cm vs.0.01 ± 0.27 cm, P = .017) and Y (0.03 ± 0.23 cm vs.0.08 ± 0.27 cm, P = .039). Irrespective of the side, mask use significantly affected the errors in all 3 directions (all P < .05). The required PTV margins (cm) were: left BCS with mask (0.51, 0.51, 0.54), and without mask (0.47, 0.49, 0.57); right BCS with mask (0.64, 0.63, 0.55), and without mask (0.59, 0.62, 0.53).

CONCLUSIONS: In SGRT-based workflows for patients with BCS, omitting the thermoplastic mask improves setup reproducibility. Right-sided BCS cases require larger margins than left-sided cases.

ADVANCES IN KNOWLEDGE: This study provides clinical recommendations for patient positioning and immobilization techniques in whole-breast radiotherapy following breast-conserving surgery.

来源:Br J Radiol,查看原文。版权归原作者所有。

引用

  1. Wan B, Han J, Zheng Y, et al. Evaluation of patient setup accuracy using optical surface guidance in postoperative radiotherapy for patients with breast cancer after breast-conserving surgery. Br J Radiol. 2026. PMID: 41904664. doi:10.1093/bjr/tqag058

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。