光子计数探测器CT的层厚对主动脉瓣钙化评分的影响
这项回顾性研究纳入84例接受心电门控光子计数探测器CT的患者,在3 mm、1 mm和0.4 mm层厚下测量主动脉瓣Agatston评分、钙体积和钙质量。层厚越薄,Agatston评分和钙体积越低,导致8%–9%的患者从CT重度重新分类为CT非重度;钙质量差异极小。
要点
- 层厚从3 mm减至0.4 mm时主动脉瓣Agatston评分和钙体积降低,钙质量几乎不变。
- 以3 mm层厚为参照,1 mm和0.4 mm层厚分别使8.3%和9.5%的患者从CT重度降为CT非重度。
- 1 mm和0.4 mm层厚对应的回归等效Agatston阈值均低于常规性别特异性阈值。
核心结论
- 薄层光子计数探测器CT降低主动脉瓣Agatston评分和钙体积,导致8%–9%的患者主动脉瓣狭窄严重程度被重新分类。
研究设计与样本
- 回顾性研究,纳入2023年3月至2024年10月接受心电门控光子计数探测器CT的患者84例(中位年龄84岁,IQR 80–88岁,女性58例),在3 mm、1 mm和0.4 mm层厚下测量主动脉瓣Agatston评分、钙体积和钙质量,采用线性混合模型和回归分析。
诊断标准与关键数值
- 采用性别特异性主动脉瓣Agatston评分阈值进行主动脉瓣狭窄严重程度分类:男性≥2000,女性≥1300。
临床应用与阅片要点
- 以3 mm层厚为参照,1 mm和0.4 mm层厚分别使7/84(8.3%)和8/84(9.5%)的患者从CT重度重新分类为CT非重度(P < .001)。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
Purpose To evaluate the effect of section thickness on aortic valve calcification quantification via photon-counting detector (PCD) CT. Materials and Methods This retrospective study included patients who underwent electrocardiography-gated PCD CT from March 2023 to October 2024. The aortic valve Agatston score (AVAS), aortic valve calcium volume (AVCV), and aortic valve calcium mass (AVCM) were measured at 3-mm, 1-mm, and 0.4-mm sections. Linear mixed model and regression analyses were performed. Sex-specific AVAS thresholds were used for aortic stenosis (AS) severity classification (≥2000 for male patients, ≥1300 for female patients). Results Eighty-four patients (median age: 84 years [IQR: 80-88 years], 58 female) were included. AVAS at 70 keV decreased with thinner sections: 3-mm section, 2141 (95% CI: 1931, 2351); 1-mm section, 1900 (95% CI: 1690, 2110), and 0.4-mm section, 1791 (95% CI: 1581, 2001) (P < .001). AVCV decreased similarly: 1806 mm3 (95% CI: 1640, 1972), 1496 mm3 (95% CI: 1330, 1661), and 1373 mm3 (95% CI: 1207, 1538) (P < .001). AVCM differed minimally: 518 mg (95% CI: 452, 585), 525 mg (95% CI: 459, 592), and 527 mg (95% CI: 461, 594) (P < .001 except for 1 mm vs 0.4 mm, P = .24). Compared with 3-mm sections, seven of 84 (8.3%) and eight of 84 (9.5%) patients were reclassified from CT-severe to CT-nonsevere with 1- and 0.4-mm sections, respectively (P < .001). Regression-based equivalent thresholds were 1770 (male patients) and 1154 (female patients) at 1 mm and 1675 (male patients) and 1087 (female patients) at 0.4 mm. Conclusion Thin-section PCD CT decreased AVAS and AVCV, resulting in AS severity reclassification in 8%-9% of patients. Keywords: Applications-CT, CT-Photon Counting, Cardiac, Aortic Valve, Calcifications/Calculi Supplemental material is available for this article. © RSNA, 2026.
引用
- Murai K, Kisohara M, Itoh T, et al. Effect of CT Section Thickness on Aortic Valve Calcium Scoring via Photon-counting Detector CT. Radiol Cardiothorac Imaging. 2026. PMID: 41885621. doi:10.1148/ryct.250221