文献速递儿科影像胸部DR

动态数字X线摄影作为微创替代灌注闪烁显像评估儿童肺血流的前瞻性研究

Dynamic digital radiography as a minimally invasive alternative to perfusion scintigraphy for assessing pulmonary blood flow in children: a prospective study.
PubMed 文献 · Pediatr Radiol · 2026-03-28

前瞻性纳入13例5–7岁先天性膈疝术后幸存儿童,同日行动态数字X线摄影(DDR)与肺灌注闪烁显像(PPS)。DDR血流比与PPS强相关(r=0.977,P<0.01),方法间绝对差<5%,无固定或比例偏倚;DDR中位有效剂量0.16 mSv,显著低于PPS的2.09 mSv。作者认为DDR可作为微创、低剂量替代方法,但需在更广泛儿科人群中进一步验证。

要点

  • DDR血流比与PPS强相关(r=0.977,P<0.01),方法间绝对差<5%。
  • DDR中位有效剂量0.16 mSv,显著低于PPS的2.09 mSv。
  • DDR无固定或比例偏倚,但仅纳入13例先天性膈疝术后儿童。

核心结论

  • DDR计算的血流比与闪烁显像结果相关性良好,可作为微创、低剂量替代PPS评估肺灌注,但需在更广泛儿科人群中进一步验证。

研究设计与样本

  • 前瞻性研究,连续纳入2024年1月至2025年6月间计划行常规PPS的5–7岁先天性膈疝幸存者13例(中位年龄6.07岁;左侧膈疝10例),同日行DDR,以PPS为参考标准。

关键影像表现/征象

  • DDR通过脉冲X线发生器与平板探测器采集序列图像,提取X线衰减细微变化以可视化和分析肺灌注。

诊断标准与关键数值

  • 计算患侧肺血流比,DDR与PPS强相关(r=0.977,P<0.01),方法间血流比绝对差<5%,无固定或比例偏倚。
  • DDR中位有效剂量0.16 mSv,显著低于PPS的2.09 mSv。

临床应用与阅片要点

  • DDR可提供与PPS一致的血流比结果且辐射剂量更低,可作为儿童肺灌注评估的微创替代选择。

局限性

  • 样本量小(13例),仅纳入5–7岁先天性膈疝术后幸存者。
  • 作者结论指出需在更广泛儿科人群中进一步验证(Future validation in a broader pediatric population is warranted)。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

BACKGROUND: In congenital diaphragmatic hernia, postoperative right-left asymmetry of pulmonary perfusion affects outcome. Although pulmonary perfusion scintigraphy (PPS) is widely used to assess pulmonary perfusion, it requires the use of radionuclides and peripheral venous access. In contrast, dynamic digital radiography (DDR) visualizes and analyzes pulmonary perfusion by extracting subtle changes in X-ray attenuation from sequential images acquired using a pulsed X-ray generator and a flat-panel detector.

OBJECTIVE: This study aimed to evaluate the accuracy of DDR for assessing pulmonary perfusion in children and verify the associated radiation dose.

MATERIALS AND METHODS: Consecutive congenital diaphragmatic hernia survivors aged 5-7 years scheduled for routine PPS between January 2024 and June 2025 underwent DDR on the same day. The images were analyzed using a dedicated workstation. For each modality, we calculated the blood-flow ratio in the affected lung and assessed the agreement using Pearson correlation and Bland-Altman analysis. Effective doses were also compared.

RESULTS: Thirteen children (median age, 6.07 years; 10 left-sided diaphragmatic hernia) were included in the study. DDR-derived blood-flow ratio strongly correlated with PPS (r=0.977, P<0.01), with absolute inter-method blood-flow ratio differences <5%. Neither fixed nor proportional bias was observed. The median effective dose for DDR was 0.16 mSv-significantly lower than that for PPS (2.09 mSv). The blood-flow ratio calculated using DDR correlated well with the scintigraphy results.

CONCLUSION: DDR could serve as a minimally invasive, low-dose alternative to PPS for the evaluation of pulmonary perfusion. Future validation in a broader pediatric population is warranted.

来源:Pediatr Radiol,查看原文。版权归原作者所有。

引用

  1. Go K, Suzuki K, Ueda K, et al. Dynamic digital radiography as a minimally invasive alternative to perfusion scintigraphy for assessing pulmonary blood flow in children: a prospective study. Pediatr Radiol. 2026. PMID: 41902931. doi:10.1007/s00247-026-06589-z

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。