文献速递骨肌介入技术与安全CTMR

利用零回波时间MRI对像差校正经脊柱聚焦超声的计算机模拟探索

An In Silico Exploration of Aberration-Corrected Transspinal Focused Ultrasound Using Zero Echo Time MRI.
PubMed 文献 · Ultrasound Med Biol · 2026-03-05

研究对11例离体胸椎进行CT和零回波时间(ZTE)MRI成像,将ZTE图像转换为伪CT,并模拟256阵元阵列的经脊柱聚焦超声,比较五种校正方法。结果显示,基于ZTE的MR校正聚焦优于未校正聚焦,但未达到CT校正的金标准水平;异质性与均匀性伪CT校正性能相似,均匀性CT校正接近金标准,提示脊柱形态的准确呈现是MR校正经脊柱聚焦超声准确性的关键。

要点

  • 基于ZTE MRI的伪CT校正可改善经脊柱聚焦超声的聚焦准确性,但未达到CT校正金标准。
  • 均匀性与异质性伪CT校正性能相似,提示准确脊柱形态表示是关键。
  • CT校正均匀性方法接近金标准,平均空间偏移0.8±1.7 mm,Dice系数0.91±0.06。

核心结论

  • 利用ZTE衍生声学特性的MR校正聚焦优于未校正聚焦,但未匹配金标准CT校正聚焦;异质性与均匀性伪CT校正性能相似及均匀性CT校正的强劲表现提示脊柱准确形态表示是MR校正经脊柱聚焦超声准确性的关键。

研究设计与样本

  • 计算机模拟研究,对11例离体胸椎进行CT和零回波时间(ZTE)MRI成像,ZTE图像通过文献线性拟合(-2085×ZTE+2329)转换为伪CT(pCT),使用256阵元阵列(400 kHz,焦点半高全宽轴向约20 mm、横向4 mm)模拟经脊柱聚焦,比较五种校正方法:未校正、pCT校正均匀特性、pCT校正异质特性、CT校正均匀特性、CT校正异质特性(金标准)。

诊断标准与关键数值

  • 异质性pCT校正与未校正相比:平均空间最大偏移1.5±2.0 mm vs. 3.5±4.1 mm,50%轮廓Sørensen-Dice相似性0.75±0.10 vs. 0.57±0.18,相对金标准的驱动效率变化-12%±11% vs. -21%±25%。
  • 均匀特性pCT校正与异质性情况性能相似。
  • CT校正均匀特性接近金标准:平均空间偏移0.8±1.7 mm,Dice系数0.91±0.06,驱动效率变化小(1%±3%)。

局限性

  • 本研究为计算机模拟(in silico)探索,使用离体胸椎标本,结果需进一步研究验证。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

This study numerically assesses the feasibility of magnetic resonance imaging (MRI)-corrected transspinal focused ultrasound (FUS). Eleven ex vivo thoracic vertebrae were imaged with computed tomography (CT) and zero echo time (ZTE) magnetic resonance imaging. ZTE images were converted to pseudo-CT (pCT) with a linear fit from the literature (-2085×ZTE+2329). Transspinal focusing was simulated with a 256-element array (400 kHz, focal full-width-at-half-maximum ∼20 mm axial, 4 mm lateral) using five correction methods: no correction; pCT-corrected, homogeneous acoustic properties; pCT-corrected, heterogeneous properties; CT-corrected, homogeneous properties; and CT-corrected, heterogeneous properties (gold standard). Pressure fields generated using each correction method were evaluated against the gold standard. Compared to no correction, heterogeneous pCT-corrected focusing reduced mean spatial maximum shift (1.5 ± 2.0 mm vs. 3.5 ± 4.1 mm), increased Sørensen-Dice similarity of the 50% contour (0.75 ± 0.10 vs. 0.57 ± 0.18), and improved the change in driving efficiency relative to the gold standard (-12% ± 11% vs. -21% ± 25%). pCT-corrected with homogeneous properties performed similarly to the heterogeneous case. CT-corrected with homogeneous properties approached the gold standard, with a mean spatial shift of 0.8 ± 1.7 mm, Dice coefficient of 0.91 ± 0.06, and little change in driving efficiency (1% ± 3%). MR-corrected focusing using ZTE-derived acoustic properties outperformed uncorrected focusing but did not match gold-standard CT-corrected focusing. Similarities in the performance of heterogeneous versus homogeneous pCT correction, along with the strong performance of homogeneous CT-corrected focusing, suggest that accurate morphological representation of the spine is key to accurate MR-corrected transspinal focused ultrasound.

来源:Ultrasound Med Biol,查看原文。版权归原作者所有。