降低剂量,保持诊断:CT引导下肺活检方案优化结果
单中心回顾性研究比较CT引导下肺活检方案优化前后各100例活检的剂量与诊断表现,优化后中位DLP从576.5降至62.0 mGycm(p<0.001),诊断敏感度、特异度、PPV不变,活检时长无显著变化。
要点
- 方案优化后中位DLP由576.5降至62.0 mGycm(p<0.001),所有操作者DLP均显著降低(p<0.01)。
- 优化前后敏感度87.2%、特异度100.0%、PPV100.0%保持不变,NPV由50.0%升至54.6%,准确率由88.7%升至88.9%。
- 活检时长优化前12.0分钟 vs 优化后14.0分钟(p=0.11),无显著变化。
核心结论
- 方案优化是降低患者辐射暴露、在CT引导下实现更安全肺活检的可行方式,且不损害诊断表现或操作时长。
研究设计与样本
- 单中心回顾性研究,剂量长度乘积、活检时长和诊断表现分析纳入200例活检(优化前100例、优化后100例);操作者特异性DLP评估另纳入22例活检,DLP趋势分析纳入221例补充活检;统计显著性设为p<0.05。
诊断标准与关键数值
- 优化前敏感度87.2%、特异度100.0%、PPV100.0%,优化后敏感度87.2%、特异度100.0%、PPV100.0%,保持不变。
- NPV由优化前50.0%升至优化后54.6%,准确率由优化前88.7%升至优化后88.9%。
- 中位DLP由优化前576.5(486.8-653.0)mGycm降至优化后62.0(53.0-73.0)mGycm(p<0.001)。
- 操作者DLP均值的标准差由169.8降至59.8 mGycm,降低64.8%。
- 活检时长优化前12.0(10.0-16.5)分钟 vs 优化后14.0(11.0-17.0)分钟(p=0.11),无显著变化。
临床应用与阅片要点
- 作者认为方案优化是降低患者辐射暴露、实现更安全CT引导下肺活检的可行方式,且不损害诊断表现或操作时长。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: To assess changes in patient radiation exposure during computed tomography (CT)-guided lung biopsies before and after protocol optimization and to evaluate the potential impact of this optimization on diagnostic yield and biopsy duration.
MATERIAL AND METHODS: In this single-center retrospective study, analysis of dose-length product (DLP), biopsy duration and diagnostic performance included data from 200 biopsies-100 performed before and 100 after protocol optimization. Evaluation of operator-specific DLP included additional data from 22 biopsies, and DLP trend analysis incorporated data from 221 supplementary biopsies. Statistical significance was set at p < 0.05.
RESULTS: Following protocol optimization, the median DLP decreased significantly from 576.5 (486.8-653.0) to 62.0 (53.0-73.0) mGycm (p < 0.001). All operators demonstrated a significant reduction in DLP (p < 0.01). Diagnostic performance was similar before and after optimization: sensitivity, specificity and PPV remained unchanged at 87.2%, 100.0% and 100.0%, respectively. NPV and accuracy increased slightly from 50.0% and 88.7% before optimization to 54.6% and 88.9% post-optimization. Standard deviation of operator DLP means decreased by 64.8% (from 169.8 to 59.8 mGycm). DLP for all operators decreased gradually before optimization, with a sharp and sustained reduction over the next 2 years. No significant change was observed in biopsy duration-12.0 (10.0-16.5) versus 14.0 (11.0-17.0) minutes (p = 0.11).
CONCLUSION: Protocol optimization is a feasible way to reduce patient radiation exposure and achieve safer lung biopsies under CT guidance without compromising diagnostic performance or procedure duration.