用于暗场胸部X线摄影个体化辐射暴露规划的多重线性回归模型
该回顾性研究分析273例在原型系统采集的后前位胸部暗场X线影像,比较基于体重、身高、年龄、性别的多重线性回归模型与仅基于BMI的简单线性回归模型对所需管电流的预测性能,结果显示普通最小二乘多重线性回归模型表现最佳(R²=0.712,RMSE=0.234,MAPE=0.033),优于BMI模型的R²=0.627、RMSE=0.266、MAPE=0.037。
要点
- 多重线性回归(体重、身高、年龄、性别)预测暗场胸片管电流优于仅用BMI的简单线性回归。
- 最佳模型为普通最小二乘多重线性回归:R²=0.712,RMSE=0.234,MAPE=0.033。
- BMI模型的拟合指标为R²=0.627,RMSE=0.266,MAPE=0.037。
核心结论
- 多重线性回归相比简单线性回归能更好地用于X射线暗场胸部摄影的个体化暴露规划。
研究设计与样本
- 回顾性研究,分析273例在原型系统采集的后前位胸部暗场X线影像,以达到目标辐射剂量所需管电流为预测目标,比较不同多重线性回归模型与BMI-based简单线性回归模型的拟合优度。
诊断标准与关键数值
- 普通最小二乘多重线性回归(基于体重、身高、年龄、性别)预测管电流:R²=0.712,RMSE=0.234,MAPE=0.033。
- BMI-based简单线性回归预测管电流:R²=0.627,RMSE=0.266,MAPE=0.037。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND: Clinical X-ray dark-field radiography has shown to be promising for visualizing different lung pathologies. To keep the radiation dose as low as reasonably achievable (ALARA principle), individualized exposure planning is necessary. However, the current scanning-based implementation of dark-field radiography complicates the use of automatic exposure control. Previously, a BMI-based linear regression model was proposed as a substitute. Here, we aim to improve this proposed model by investigating multiple linear regression for patient-individual exposure planning of dark-field chest radiography.
METHODS: For this retrospective study, 273 posteroanterior thorax images acquired at a prototype system for dark-field chest radiography were analyzed retrospectively regarding the X-ray tube current needed to achieve the target radiation dose. Different multiple linear regression models were tested to find the optimal multiple regression model for predicting the necessary tube current based on a person's weight, height, age, and sex. R2 score, the root mean square error (RMSE), and the mean absolute percentage error (MAPE) were used to evaluate the goodness-of-fit of different regression models. Each model was also compared to a BMI-based model.
RESULTS: To predict the target tube current for dark-field chest radiography, multiple linear regression using ordinary least squares performed best (R2 = 0.712, RMSE = 0.234, and MAPE = 0.033). In comparison, simple linear regression using only the body mass index achieved only R2 = 0.627, RMSE = 0.266, and MAPE = 0.037.
CONCLUSION: Multiple linear regression allows better exposure planning in X-ray dark-field chest radiography than simple linear regression.
引用
- Bast H, Dorosti T, Lochschmidt ME, et al. Multiple linear regression models for individualized radiation exposure planning in dark-field chest radiography. Eur J Radiol. 2026. PMID: 41806465. doi:10.1016/j.ejrad.2026.112765
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。