MARS MRI 诊断非骨水泥全髋关节置换术后无菌性股骨柄松动
回顾性纳入114例全髋关节翻修术前行MARS MRI的病例,两名盲法肌肉骨骼放射科医生按14个Gruen分区评估假体周围骨吸收、骨髓水肿、骨膜反应和假体周围骨溶解,以翻修术中所见为金标准。预测模型(近端骨吸收、中远端骨吸收、中远端骨膜反应、远端骨髓水肿)诊断股骨柄松动的敏感度0.708、特异度0.970、阳性预测值0.944、阴性预测值0.821;骨吸收、骨髓水肿和骨膜反应在固定柄近端也可出现,而在中远端柄区提示松动。
要点
- 中远端Gruen区的假体周围骨吸收和骨膜反应提示股骨柄松动,近端出现也可见于固定柄。
- 预测模型敏感度0.708、特异度0.970、阳性预测值0.944、阴性预测值0.821。
- 观察者间一致性Gwet AC1:骨吸收0.80–0.98、骨髓水肿0.91–0.99、骨膜反应0.87–0.97。
核心结论
- MARS MRI评估股骨柄松动具有可重复性和准确性;假体周围骨吸收、骨髓水肿和骨膜反应在固定全髋关节置换近端区域也可出现,而在中远端柄区提示松动。
研究设计与样本
- 回顾性纳入114例连续病例,均为全髋关节置换翻修术前接受髋部MARS MRI者;两名盲法肌肉骨骼放射科医生独立按14个Gruen分区评估;以翻修术中所见为金标准;采用二项逻辑回归建立预测模型以最大化阳性预测值和准确性;观察者间信度采用绝对一致率、Cohen κ和Gwet AC1评估。
关键影像表现/征象
- 除Gruen 11区外,松动柄在所有Gruen区的假体周围骨吸收发生率均显著高于固定柄。
- 固定柄近端也可见假体周围骨吸收(最高达23%),而中远端骨吸收罕见(≤3%)。
- 骨髓水肿在所有柄的近端最普遍(固定柄/松动柄:39%/60%),内侧和远端差异显著。
- 骨膜反应在松动柄的中远端Gruen区显著更常见。
- 假体周围骨溶解罕见,最多见于Gruen 7区。
诊断标准与关键数值
- 预测模型纳入近端假体周围骨吸收、中远端假体周围骨吸收、中远端骨膜反应和远端骨髓水肿,诊断股骨柄松动的敏感度0.708、特异度0.970、阳性预测值0.944、阴性预测值0.821;观察者间一致性Gwet AC1在评估区域为骨吸收0.80–0.98、骨髓水肿0.91–0.99、骨膜反应0.87–0.97。
临床应用与阅片要点
- 阅片时按Gruen分区评估假体周围骨吸收、骨髓水肿和骨膜反应;近端上述征象也可见于固定柄,而中远端柄区出现则提示松动;假体周围骨溶解罕见,最多见于Gruen 7区。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND: Despite the increasing use of MARS (metal artifact reduction sequence) MRI to investigate painful total hip arthroplasties (THA), no validated criteria exist for diagnosing femoral stem loosening.
PURPOSE: To evaluate MARS MRI for the diagnosis of aseptic stem loosening and determine its diagnostic accuracy.
MATERIALS AND METHODS: One hundred fourteen consecutive cases with THA revision surgery and MARS MRI of the hip were retrospectively included. Two blinded musculoskeletal radiologists independently assessed periprosthetic bone resorption (PPBR), bone marrow edema (BME), periosteal reaction (PR), and periprosthetic osteolysis (PO) in 14 Gruen zones (GZ). Intraoperative findings at revision surgery served as the ground truth. A predictive model was created using binomial logistic regression models to predict the probability of a loose stem with maximizing positive predictive value (PPV) and accuracy. Interobserver reliability was assessed with absolute agreement, Cohen κ and Gwet AC1.
RESULTS: During surgery, 66 stems were fixed and 48 loose. PPBR occurred significantly more frequently in loose stems across all GZs except GZ11. Proximal PPBR was also observed in fixed stems (up to 23%), whereas middle and distal PPBR were rare (≤3%). BME was most prevalent proximally in all stems (fixed/loose: 39%/60%) with significant differences medially and distally. PR was significantly more frequent in loose stems in the middle and distal GZs. PO were rare, most occurred in GZ7. The predictive model considering proximal PPBR, mid-distal PPBR, mid-distal PR, and distal BME performed with a sensitivity of 0.708, specificity of 0.970, PPV 0.944, negative predictive value 0.821. Interobserver agreement (Gwet AC1) in the considered zones was for PPBR between 0.80 and 0.98, BME 0.91 to 0.99, PR 0.87 to 0.97.
CONCLUSIONS: MARS MRI is reproducible and accurate for assessing stem loosening. PPBR, BME, and PR can also be found in fixed THA in the proximal region, whereas they indicate loosening in the middle and distal stem region.
引用
- Aepli M, Fischer A, Rüdiger HA, et al. MARS MRI for the Diagnosis of Aseptic Stem Loosening in Cementless Total Hip Arthroplasty. Invest Radiol. 2026. PMID: 41542796. doi:10.1097/RLI.0000000000001256
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。