窄带脉冲定量多频超声成像用于组织表征
本研究实现了一种窄带脉冲多频超声(MFUS)采集序列,并在体模和代谢功能障碍相关脂肪性肝病(MASLD)小鼠模型中验证其定量测量。结果表明MFUS对衰减系数和散射体尺寸变化敏感,临床前结果初步验证了MFUS参数用于肝脏组织表征的性能。
要点
- MFUS通过频移窄带脉冲获取背散射数据,可同时估算衰减系数和相对散射体尺寸。
- 体模中实测衰减系数与真实值接近,MFUS较B超提高异质区域对比噪声比。
- MCD饮食小鼠肝脏衰减系数升高、相对散射体尺寸降低,与组织学一致。
核心结论
- MFUS成像对体模衰减系数和散射体尺寸变化敏感,临床前结果初步验证MFUS参数用于肝脏组织表征的性能,衰减系数和相对散射体尺寸变化与既往研究一致。
研究设计与样本
- 临床前方法学研究,使用可编程研究扫描仪(Vantage NXT 256,Verasonics Inc)和L11-5v线阵探头。
关键影像表现/征象
- MFUS梯度定义为对数B超强度与发射频率之间的线性斜率,衰减系数和相对散射体尺寸由该梯度参数导出。
- 异质体模中B超与MFUS估算相对散射体尺寸的平均对比噪声比分别为0.11和0.30,MFUS较高。
- MCD饮食小鼠肝脏表现为衰减系数升高、相对散射体尺寸降低,与对照动物相比差异有统计学意义(P < 0.001)。
诊断标准与关键数值
- 双层层状体模真实衰减值分别为-0.70和-0.95 dB/cm/MHz,MFUS估计值分别为-0.71 ± 0.04和-0.93 ± 0.03 dB/cm/MHz。
- 均匀体模嵌入38至89 µm散射体,不同尺寸散射体的相对散射体尺寸估计值差异有统计学意义(P < 0.001)。
局限性
- 原文指出未来研究将进一步探索优化和潜在应用。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: Multifrequency ultrasound (MFUS) is a novel quantitative imaging technique that captures backscattered data after transmission of a sequence of frequency-shifted narrowband ultrasound (US) pulses. The goal of this study was to implement a MFUS acquisition sequence and then validate quantitative measurements using an animal model of metabolic dysfunction-associated steatotic liver disease (MASLD).
MATERIALS AND METHODS: US imaging was performed using a programmable research scanner (Vantage NXT 256, Verasonics Inc) equipped with a L11-5v linear array. Radiofrequency (RF) data were acquired from 24 different frequency-shifted narrowband transmit pulses using plane wave imaging with spatial angular compounding. The MFUS gradient was defined as the linear slope between logarithmic B-scan intensity and transmit frequency. The US attenuation coefficient and relative scatterer size (RSS) were derived from the MFUS gradient parameter. A bilayer tissue-mimicking phantom with regions of known US attenuation values of -0.70 and -0.95 dB/cm/MHz, respectively, was used for in vitro MFUS imaging and attenuation coefficient estimation. Custom homogeneous phantoms containing 38 to 89 µm-sized US scatterers were also used to compare RSS estimates. A heterogeneous phantom containing a circular target and background material embedded with 64 and 38 µm-sized scatterers, respectively, was used to measure the contrast-to-noise ratio (CNR) between the two regions for B-scan and MFUS-derived RSS estimates. Preclinical US imaging was performed on mice fed standard chow or a methionine and choline-deficient (MCD) diet to induce liver steatosis ( n =5 animals per group). Mice underwent in vivo MFUS imaging, and then attenuation coefficient and RSS estimates were derived for each mouse liver.
RESULTS: In vitro attenuation estimates were found to be -0.71 ± 0.04 and -0.93 ± 0.03 dB/cm/MHz for a phantom with true attenuation values of -0.70 and -0.95 dB/cm/MHz, respectively. A difference was found in RSS values from homogenous phantoms embedded with different-sized US scatterers and CNR measurements from the heterogeneous inclusion phantom ( P < 0.001). Average CNR values increased from 0.11 to 0.30 when using B-scan and MFUS imaging, respectively. Relative to control animals, in vivo liver imaging studies revealed increased attenuation coefficient and decreased RSS values in mice fed the MCD diet ( P < 0.001). After US imaging, all animals were humanely euthanized and livers excised for histologic analysis.
CONCLUSION: MFUS imaging of phantom materials revealed a sensitivity to changes in attenuation coefficient and US scatterer size conditions. Preliminary preclinical results validated the performance of MFUS-derived parameters for liver tissue characterization. Changes in attenuation coefficient and RSS values were consistent with previous studies. Overall, MFUS is an innovative imaging technique, and future studies will explore further optimization and potential applications.