单能重建与超高分辨率对光子计数探测器CT检测小内漏的影响

Effect of Monoenergetic Reconstructions and Ultra-High-Resolution on the Detection of Small Endoleaks With Photon-counting Detector CT.
PubMed 文献 · Invest Radiol · 2026-09-01

研究在模拟中型和大型患者的动脉瘤体模上用光子计数探测器CT扫描,比较不同keV单能图像及超高分辨率重建对检测小内漏的影响。三名放射科医师盲法评估含2880个模拟内漏的80组CT序列,结果显示55 keV标准或超高分辨率及60 keV标准重建检测小内漏表现最佳,超高分辨率在50 keV时对4至6 mm及6 mm内漏有提升。

要点

  • 在实验体模中,55 keV标准或超高分辨率及60 keV标准重建检测小内漏表现最佳。
  • 50 keV时超高分辨率对4至6 mm内漏的A1值优于标准重建(中型体模0.82对0.72,P=0.015)。
  • 阅片者表现与图像噪声负相关,与内漏对比噪声比无相关性。

核心结论

  • 55 keV标准或超高分辨率及60 keV标准重建在PCD-CT检测小内漏中表现最佳,超高分辨率在50 keV时对4至6 mm和6 mm内漏有提升。

研究设计与样本

  • 在模拟中型和大型患者的动脉瘤体模上用光子计数探测器CT扫描,单能图像以标准分辨率40、45、50、55、60 keV及超高分辨率45、50、55 keV重建,三名独立盲法放射科医师评估含2880个直径2、4、6 mm模拟内漏的80组CT序列,采用JAFROC1方法计算A1值并与客观图像质量做Spearman相关分析。

诊断标准与关键数值

  • 中型体模:50 keV超高分辨率、55 keV标准、55 keV超高分辨率及60 keV标准序列阅片者A1=0.66至0.72。
  • 大型体模:55 keV标准、55 keV超高分辨率及60 keV标准重建阅片者A1=0.40至0.49。
  • 50 keV时超高分辨率对4至6 mm内漏A1:中型体模0.82对0.72(P=0.015),大型体模0.42对0.30(P=0.028)。
  • 50 keV时超高分辨率对6 mm内漏A1:中型体模0.93对0.83(P=0.002),大型体模0.53对0.36(P=0.013)。
  • 阅片者表现与图像噪声负相关:中型体模Spearman R=-0.83(P=0.01),大型体模Spearman R=-0.98(P<0.001);与内漏对比噪声比无相关性(P=0.91和0.73)。

临床应用与阅片要点

  • 在实验体模中,55 keV标准或超高分辨率及60 keV标准重建检测小内漏表现最佳。
  • 50 keV时超高分辨率对4至6 mm和6 mm内漏的检测表现优于标准重建。

局限性

  • 本研究为实验体模研究,结果基于模拟内漏和体模模型。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

BACKGROUND AND OBJECTIVES: Following endovascular repair of an aortic aneurysm, patients need regular surveillance using CT to exclude endoleaks. The optimal technical parameters for photon-counting detector CT (PCD-CT) have not yet been established in this clinical context. This study examines the impact of the ultra-high resolution (UHR) mode and monoenergetic reconstructions at various keV on the detection of small endoleaks in experimental settings.

MATERIALS AND METHODS: An aneurysm phantom model mimicking medium- and large-sized patients was scanned using PCD-CT. Monoenergetic images (MEIs) were reconstructed with standard resolution at 40, 45, 50, 55, and 60 keV and with UHR at 45, 50, and 55 keV. Three independent blinded radiologists assessed 80 CT series containing 2880 simulated endoleaks with diameters of 2, 4, and 6 mm. The observers' ratings were analyzed and compared with the jackknife alternative free-response operational characteristics (JAFROC1) method by calculating a figure-of-merit for the performance ( A1 ). The Spearman R was calculated for the correlation between A1 and objective image quality.

RESULTS: The best readers' performance was achieved with the 50 keV UHR, 55 keV standard, 55 keV UHR, and 60 keV standard series ( A1 =0.66 to 0.72) in the medium and with the 55 keV standard, 55 keV UHR, and 60 keV standard reconstructions ( A1 = 0.40 to 0.49) in the large phantom. UHR provided higher A1 than the standard reconstruction at 50 keV for 4 to 6 mm endoleaks ( A1 , 0.82 vs. 0.72; P =0.015 in medium and 0.42 vs. 0.30; P =0.028 in large phantom) and for 6 mm endoleaks ( A1 , 0.93 vs. 0.83; P =0.002 in medium and 0.53 vs. 0.36; P =0.013 in large phantom). In both phantoms, readers' performance correlated negatively with image noise (Spearman R =-0.83; P =0.01 in the medium and Spearman R =-0.98; P <0.001 in the large phantom) and showed no correlation with the contrast-to-noise ratio of endoleaks ( P =0.91 and 0.73).

CONCLUSIONS: The 55 keV standard or UHR and 60 keV standard reconstructions yielded the best performance for detecting small endoleaks with PCD-CT. The UHR mode improved performance at 50 keV for 4 to 6 mm and 6 mm endoleaks.

来源:Invest Radiol,查看原文。版权归原作者所有。