文献速递骨肌技术与安全CT

负载钆纳米颗粒的可降解骨水泥的光子计数CT成像

Photon-counting Computed Tomography of Degradable Bone Cement Loaded With Gadolinium Nanoparticles.
PubMed 文献 · Invest Radiol · 2026-09-01

本研究将钆纳米颗粒(GdNP)加入可吸收骨水泥以提高其射线不透性,并用临床前micro-EID-CT和micro-PCCT评估骨水泥与椎骨的对比噪声比(CNR),同时评估PCCT材料分解定量钆浓度的准确性。结果显示PCCT的CNR总体高于EID-CT,材料分解与GdNP质量分数高度相关(r=0.9753),钆定量均方根误差为1.60 mg/mL。

要点

  • PCCT图像CNR总体高于EID-CT,42-51 keV能量箱CNR最大。
  • PCCT材料分解可定量钆浓度,与标称浓度高度相关(r=0.999)。
  • 负载GdNP骨水泥与椎骨的CNR随GdNP质量分数增加而升高。

核心结论

  • 加入GdNP可提高骨水泥与椎骨的CNR,PCCT的CNR总体高于EID-CT,材料分解可准确量化溶液中的钆浓度并改善负载GdNP骨水泥与钙基椎体的视觉区分,这些方法可能用于改善植入骨水泥的监测。

研究设计与样本

  • 临床前研究:合成GdNP并表征,将羟基磷灰石基骨水泥负载0%至10% w/w的GdNP,用临床前micro-EID-CT和micro-PCCT扫描评估CNR,并用钆材料分解图像测量钆含量,另对0至20 mg/mL钆标准品行micro-PCCT成像。

关键影像表现/征象

  • 合成的GdNP平均直径为15.42±1.82 nm。
  • EID-CT和PCCT的信号强度均随GdNP质量分数增加而升高。
  • PCCT各能量箱中,8%与10% GdNP负载骨水泥的CNR均与无GdNP负载骨水泥显著不同(P<0.05)。
  • 42-51 keV能量箱的CNR总体大于其他能量箱。
  • PCCT图像CNR总体大于EID-CT。

诊断标准与关键数值

  • EID-CT图像中,GdNP负载≥4%的骨水泥相对于骨的CNR高于无GdNP负载骨水泥(P<0.05)。
  • PCCT材料分解图像测得的骨水泥钆浓度与GdNP质量分数相关(r=0.9753)。
  • PCCT估计的钆浓度与钆标准品标称浓度高度相关(r=0.999),均方根误差为1.60 mg/mL。

临床应用与阅片要点

  • PCCT材料分解可定量钆浓度并改善负载GdNP骨水泥与钙基椎体的视觉区分。
  • 42-51 keV能量箱的CNR最大,阅片时可关注该能量范围。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: This study aims to improve the radiopacity of absorbable bone cements through the addition of gadolinium nanoparticles (GdNP). We also aim to determine whether photon-counting CT (PCCT) provides superior contrast-to-noise ratio (CNR) between GdNP-loaded bone cement and vertebral bone when compared with energy-integrating CT (EID-CT), and to evaluate the accuracy of PCCT material decomposition for quantifying gadolinium concentration in solution and the GdNP-loaded cements.

MATERIALS AND METHODS: GdNPs were synthesized using a one-pot thermal decomposition method and characterized using transmission electron microscopy and dynamic light scattering. Hydroxyapatite-based bone cement was loaded with varying mass fractions of GdNPs (0% to 10% w/w), and the CNR between the GdNP-loaded cement and vertebral bone was evaluated using preclinical micro-EID-CT and micro-PCCT scanners. Gadolinium material decomposition images were used to measure the amount of gadolinium present in each of the cements. In addition, gadolinium standards (0 to 20 mg/mL) were imaged with a preclinical micro-PCCT, and the concentration of gadolinium in the vials was estimated using gadolinium material decomposition images.

RESULTS: The synthesized GdNPs had a mean diameter of 15.42±1.82 nm. Signal intensity increased with increasing mass fractions of GdNPs for both EID-CT and PCCT. In EID-CT images, cements with ≥4% GdNP loading had higher CNRs relative to bone than the cement with no GdNP loading ( P <0.05). The CNR between the 8% and 10% GdNP-loaded bone cement significantly differed from than the bone cement with no GdNP loading for all PCCT energy bins ( P <0.05). The 42-51 keV energy bin yielded the largest CNRs overall when compared with the CNRs of other energy bins. Overall, the CNRs obtained from PCCT images were larger than the EID-CT CNRs. The concentration of gadolinium in the cements measured using the PCCT material decomposition images was correlated with the mass fractions of GdNP ( r =0.9753). Estimated gadolinium concentrations were highly correlated with the nominal concentration of the gadolinium standards ( r =0.999) and the PCCT was able to accurately quantify gadolinium concentrations with a root mean square error of 1.60 mg/mL.

CONCLUSIONS: The use of GdNPs led to a higher cement-vertebra CNR for both EID-CT and PCCT. Overall, PCCT demonstrated higher CNRs than EID-CT. Material decomposition successfully quantified the concentration of gadolinium in vials and allowed for improved visual differentiation of the GdNP - loaded bone cement from the calcium-based vertebral bodies. Thus, the incorporation of radiopaque GdNPs and imaging with PCCT improved visualization of the bone cement. These methods could be used to improve monitoring of implanted bone cements. In addition, PCCT material decomposition enabled accurate quantification of gadolinium in solution.

来源:Invest Radiol,查看原文。版权归原作者所有。

引用

  1. Marks AE, Marco KPD, Barcena AJR, et al. Photon-counting Computed Tomography of Degradable Bone Cement Loaded With Gadolinium Nanoparticles. Invest Radiol. 2026. PMID: 41416961. doi:10.1097/RLI.0000000000001267

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。