斑点追踪超声评估肩峰下疼痛综合征患者肩袖肌肉功能的信度与一致性

Reliability and Agreement of Speckle-Tracking Ultrasonography for Assessing Muscle Function of the Rotator Cuff in Patients with Subacromial Pain Syndrome.
PubMed 文献 · Ultrasound Med Biol · 2025-12-13

本研究评估斑点追踪超声(STU)后处理流程的评定者间信度与一致性,以及同一评定者评估肩峰下疼痛综合征(SAPS)患者冈上肌和冈下肌功能的组内重测信度与一致性。结果显示SAPS患者评定者间信度良好至优秀,但重测信度差至中等,作者认为可能与症状相关变异有关。

要点

  • SAPS患者肩部STU后处理应变测量的评定者间信度为良好至优秀(ICC 0.87-0.97)。
  • SAPS患者组内重测信度差至中等(ICC 0.16-0.59),MDC为3.5-8.9%应变。
  • 对照组组内重测信度为良好至优秀(ICC 0.83-0.96),MDC为1.8-3%应变。

核心结论

  • STU对冈上肌和冈下肌应变测量显示出优秀的评定者间信度,但SAPS患者组内重测信度低,可能归因于症状相关变异;STU在评估肌肉功能方面有前景,但其临床适用性和可行性尚未探索。

研究设计与样本

  • 研究纳入24例SAPS患者评估评定者间和组内重测(间隔1周)变异,另纳入22例年龄和性别匹配的对照进行组内重测分析;STU记录使用EchoPAC软件后处理;信度用组内相关系数(ICC)和最小可检测变化(MDC)评估,一致性用Bland-Altman图评估。

诊断标准与关键数值

  • SAPS患者肩部评定者间信度:ICC 0.87-0.97,MDC 1.4-2.2%应变。
  • SAPS患者肩部组内重测信度:ICC 0.16-0.59,MDC 3.5-8.9%应变。
  • 对照组组内重测信度:ICC 0.83-0.96,MDC 1.8-3%应变。
  • SAPS患者肩部评定者间一致性高,对照组组内重测一致性中等,SAPS患者肩部组内重测一致性低。

临床应用与阅片要点

  • 作者认为STU在评估肌肉功能方面有前景,但其临床适用性和可行性尚未探索。

局限性

  • STU的临床适用性和可行性尚未探索。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

Shoulder pain mainly originates from subacromial pain syndrome (SAPS), with the supraspinatus (SS) and infraspinatus (IS) muscles most often affected. Speckle-tracking ultrasonography (STU) is a non-invasive method to quantify muscle deformation. However, reliability and agreement with software post-processing procedures (inter-rater) and symptom-related variance (test-retest) for the same rater are unknown. This study aimed to assess the inter-rater reliability and agreement of STU post-processing procedures and the test-retest reliability and agreement of STU for evaluating SS and IS muscle function in patients with SAPS and a healthy control group METHODS: Twenty-four SAPS patients were assessed for inter-rater and test-retest (+1 week) variability, with the addition of 22 matched (age/gender) controls for the test-retest analysis. STU recordings were post-processed using EchoPAC software. Reliability was assessed using the intra-class correlation coefficient (ICC) and minimal detectable change (MDC). Agreement was evaluated using Bland-Altman plots RESULTS: Inter-rater reliability for SAPS shoulders was good-excellent (ranges: ICC 0.87-0.97; MDC 1.4-2.2% strain) and test-retest poor-moderate (ranges: ICC 0.16-0.59; MDC 3.5-8.9% strain). For the controls, test-retest showed good-excellent reliability (ranges: ICC 0.83-0.96; MDC 1.8-3% strain). Agreement was high for SAPS shoulders between raters. Test-retest agreement was moderate for the controls and low for SAPS shoulders CONCLUSION: STU showed excellent inter-rater reliability for post-processing strain measurements of the SS and IS muscles. However, test-retest reliability was low in SAPS patients, likely due to symptom-related variability. While STU shows promise for assessing muscle function, its clinical applicability and feasibility remain unexplored.

来源:Ultrasound Med Biol,查看原文。版权归原作者所有。