将先进ClearIQ引擎(AiCE)深度学习重建算法引入临床放疗工作流程
本研究通过体模与临床扫描评估深度学习重建算法AiCE的图像质量及其对放疗计划的影响,并与AIDR3D重建比较。体模中两者水的CT值差异为5–6 HU,在IPEM 91容差范围内;临床计划剂量最大差异为0.45 Gy,中位差异为0.02 Gy。据此优化乳腺放疗CT协议后,图像质量临床可接受,平均DLP降低12%。
要点
- 体模中AIDR3D与AiCE图像水的CT值差异为5–6 HU,在IPEM 91容差范围内。
- 临床计划剂量差异最大为0.45 Gy,中位差异为0.02 Gy。
- 引入AiCE优化乳腺放疗CT协议后,平均DLP降低12%,图像质量临床可接受。
核心结论
- AiCE的CT值稳定性适合用于放疗治疗计划,引入AiCE后乳腺CT协议得以优化,在未影响图像质量的情况下降低了剂量。
研究设计与样本
- 本研究通过体模测试与临床扫描评估AiCE重建,体模为电子密度体模,采用临床协议在不同mAs下分别行AIDR3D与AiCE重建。
- 对临床扫描采用AiCE重建并重新计算临床计划的剂量分布。
- 体模测试证明降低剂量时图像质量稳定后,启动了针对乳腺放疗患者的优化项目。
关键影像表现/征象
- 电子密度体模中,AIDR3D与AiCE图像水的CT值差异为5–6 HU,具体取决于mAs。
诊断标准与关键数值
- 体模中AIDR3D与AiCE水的CT值差异为5–6 HU,在IPEM 91的HU变化容差范围内。
- 比较计划计算的临床目标时,剂量差异最大为0.45 Gy,中位剂量差异为0.02 Gy。
- 乳腺CT协议优化后平均DLP降低12%,图像质量临床可接受。
临床应用与阅片要点
- 作者通过体模与临床计划剂量评估,认为AiCE的CT值稳定性适合用于放疗治疗计划。
- 作者在乳腺放疗患者中引入AiCE优化CT协议,在图像质量临床可接受的前提下降低了剂量。
局限性
- 原文未报告局限性
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: This work aimed to introduce Advanced intelligence Clear-IQ Engine (Canon, Japan) (AiCE) reconstruction to the clinical workflow by assessing the image quality and impact on radiotherapy treatment planning. This introduction would then allow for optimization of CT scanning doses.
METHODS: A CT scan of an electron density phantom was performed using clinical protocols using AIDR3D and AiCE reconstructions at different mAs values. Clinical scans were reconstructed using AiCE and the dose distributions of the clinical plans recalculated. An optimization project was initiated for breast radiotherapy patients after the phantom testing proved the image quality is stable with reducing dose.
RESULTS: In the electron density phantom the difference between the Hounsfield Number of water for AIDR3D and AiCE images was 5-6 HU depending on the mAs. This was within IPEM 91 tolerances for HU variation. When comparing the clinical goals of plans calculated the dose difference was a maximum of 0.45 Gy. The median dose difference was a 0.02 Gy difference. The clinical doses were successfully reduced to the limit of what was tested in phantom measurements with the image quality being clinically acceptable and a 12% reduction in mean DLP.
CONCLUSION: AiCE has been evaluated to demonstrate HU stability that is suitable for use in Radiotherapy treatment planning. The breast CT protocol was optimized with the introduction of AiCE and the doses were reduced without affecting image quality.
ADVANCES IN KNOWLEDGE: AiCE is a novel reconstruction technique and has been used for clinical treatment planning and for optimizing CT doses.