乳腺活检标记物彩色多普勒闪烁伪像的水静压过压与微型CT成像分析

Analysis of Color Doppler Twinkling on Breast Biopsy Markers With Hydrostatic Overpressurization and Micro-CT Imaging.
PubMed 文献 · Ultrasound Med Biol · 2025-12-10

研究在压力舱中将闪烁金属临床标记物与闪烁聚甲基丙烯酸甲酯(PMMA)标记物周围水静压从0升至1000 psi,并用微型CT分析内部空隙与表面粗糙度;结果显示闪烁标记物随水静压升高闪烁减弱,内部空隙与闪烁无相关性,表面粗糙度参数(Sa和y维半高全宽)与闪烁相关(p=0.037和p=0.011),支持表面振荡微气泡为闪烁成因。

要点

  • 闪烁标记物随水静压升高闪烁信号减弱,支持表面微气泡振荡机制。
  • 内部空隙与闪烁无相关性,表面粗糙度参数Sa和y维半高全宽与闪烁相关。
  • PMMA标记物在聚合反应约75%硬化时间混合者高压下闪烁信号最高。

核心结论

  • 结果支持标记物表面振荡微气泡是彩色多普勒闪烁伪像的成因,较粗糙表面存在的微气泡印证了既往关于表面粗糙度与闪烁相关的发现。

研究设计与样本

  • 体外实验研究,将闪烁金属临床标记物和闪烁PMMA标记物置于压力舱,水静压从0升至1000 psi,并用微型CT扫描分析内部空隙和表面粗糙度,另测试低闪烁PMMA样本和不闪烁金属棒。

关键影像表现/征象

  • 闪烁标记物随水静压升高闪烁减弱。
  • PMMA标记物在聚合反应约75%硬化时间混合者高压下闪烁信号最高。
  • 内部空隙与闪烁无相关性。
  • 表面粗糙度参数Sa和y维半高全宽较高者与闪烁相关(p=0.037和p=0.011)。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: Previous development of a breast biopsy marker that can be readily detected using color Doppler twinkling has been reported, but the cause of twinkling remains unclear. The objective is to better understand the mechanism for twinkling by exploring the effects of microbubbles on the surfaces of markers that twinkle compared to those that do not.

METHODS: Twinkling metallic clinical markers and twinkling polymethyl methacrylate (PMMA) markers were evaluated in a pressure chamber. The hydrostatic pressure around the markers was increased from 0 to 1000 psi. To evaluate the effects of internal and external structures on twinkling, micro-CT scans were conducted to analyze the internal voids and surface roughness of the markers. A low-twinkling PMMA sample and a non-twinkling metal rod were also tested.

RESULTS: Twinkling markers exhibited decreased twinkling as the hydrostatic pressure increased. PMMA markers mixed for approximately 75% of the time required to reach the hardening phase of the polymerization reaction exhibited the highest twinkling signals under increased pressure. Internal voids were found to have no correlation with twinkling. Samples with higher surface roughness values (Sa and full width at half maximum in the y-dimension) were found to correlate with twinkling (p = 0.037 and p = 0.011, respectively).

CONCLUSION: These results support oscillating microbubbles on surfaces as a cause of twinkling. Prior work connecting surface roughness to twinkling is, hence, corroborated by the microbubbles present on rougher surfaces.