高管电流X射线管对低峰值千伏(kVp)增强胸腹盆CT中CT剂量指数容积的影响

Effect of high-tube current x-ray tube on computed tomography (CT) dose-index volume in low kilovolt peak (kVp) contrast-enhanced chest-abdomen-pelvis computed tomography (CT).
PubMed 文献 · Clin Radiol · 2025-11-21

研究前瞻性随机将118、104、100、106例受试者分入四组(A组120 kVp/835 mA、B组80 kVp/700 mA、C组120 kVp/900 mA、D组80 kVp/1,300 mA),比较体重与CTDIvol的关系。B组在体重大于57 kg时CTDIvol低于A组,D组在体重大于68 kg时CTDIvol低于C组;B组在57-67 kg时腹盆部背景噪声高于A组,C组与D组间无差异。结论认为高管电流输出CT扫描仪需将低kVp扫描的体重阈值从57 kg提高至68 kg以实现CTDIvol降低。

要点

  • 低kVp扫描降低CTDIvol的体重阈值随管电流输出能力提高而升高:旧扫描仪为57 kg,新高管电流扫描仪为68 kg。
  • 80 kVp/700 mA组在体重>57 kg时CTDIvol低于120 kVp/835 mA组(P<.001-.004)。
  • 80 kVp/1,300 mA组在体重>68 kg时CTDIvol低于120 kVp/900 mA组(P<.001-.04)。

核心结论

  • 高管电流输出CT扫描仪需将低kVp扫描的体重阈值从57 kg提高至68 kg以实现CTDIvol降低。

研究设计与样本

  • 前瞻性随机研究,2021年9月至12月行增强胸腹盆CT的受试者按kVp和最大管电流随机分为四组:A组(120 kVp,835 mA,n=118)、B组(80 kVp,700 mA,n=104)、C组(120 kVp,900 mA,n=100)、D组(80 kVp,1,300 mA,n=106),比较体重与CTDIvol的关系用非线性回归分析,并比较A vs B和C vs D的背景噪声。

关键影像表现/征象

  • B组在体重大于57 kg时CTDIvol低于A组(P<.001-.004)。
  • D组在体重大于68 kg时CTDIvol低于C组(P<.001-.04)。
  • B组在体重57-67 kg受试者腹盆部背景噪声高于A组(P<.001 for each)。
  • C组与D组背景噪声无差异(P=.21-.82)。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

AIM: The aim of this study is to investigate the relationship between participants' body weights and radiation doses in contrast-enhanced chest-abdomen-pelvis computed tomography (CT) scans, using both older and newer CT scanners that can provide adequate output even at low tube voltages.

MATERIALS AND METHODS: Participants who underwent contrast-enhanced chest-abdomen-pelvis CT from September to December 2021 were prospectively randomised into four groups based on kilovolt peak (kVp) and maximum tube current on two scanners-group A (120 kVp, 835 mA), group B (80 kVp, 700 mA), group C (120 kVp, 900 mA), and group D (80 kVp, 1,300 mA). The relationships between the participants' body weights and CT dose-index volume (CTDIvol) were compared among the four groups using non-linear regression analysis. The background noise was compared between groups A vs B and C vs D.

RESULTS: A total of 118, 104, 100, and 106 participants were included in groups A, B, C, and D, respectively. The CTDIvol was lower in group B than in group A above the body weight of 57 kg (P < .001-.004). Similarly, the CTDIvol was lower in group D than in group C above a body weight of 68 kg (P < .001-.04). The background noise was higher in group B than in group A at abdominal and pelvic regions for participants weighing 57-67 kg (P < .001 for each); however, no difference was found between groups C and D (P = .21-.82).

CONCLUSION: High-tube current output CT scanners necessitate an increase in the participants' body weight threshold for low-kVp scans from 57 kg to 68 kg to achieve a reduction in CTDIvol.

来源:Clin Radiol,查看原文。版权归原作者所有。