PubMed 文献 · Eur J Radiol · 2025-11-24

纳入40例临床怀疑骨侵犯的头颈部肿瘤患者,均使用15通道牙科线圈与标准20通道线圈行MRI,两名放射科医生回顾性评估信噪比、对比噪声比及骨侵犯等主观指标,并与组织病理学对照。15通道线圈肿瘤SNR与CNR更高,图像质量及骨侵犯评估更优,检测骨侵犯的敏感度(100% vs 76%)与特异度(97% vs 82%)均高于标准线圈。

要点

  • 15通道牙科线圈肿瘤SNR 12.1 vs 标准20通道线圈9.9,CNR 3.6 vs 1.5
  • 15通道线圈检测骨侵犯敏感度100% vs 76%,特异度97% vs 82%
  • 15通道线圈在图像质量、金属伪影、病灶勾画及骨侵犯评估方面均优于标准线圈

核心结论

  • 与标准20通道线圈相比,15通道牙科线圈MRI在头颈部肿瘤评估中提供显著更好的图像质量和诊断准确性,尤其在骨侵犯病例中。
  • 15通道线圈检测骨侵犯的敏感度(100% vs 76%)和特异度(97% vs 82%)均高于标准线圈。
  • 23例组织病理学证实骨侵犯的肿瘤中,15通道线圈正确检出23例,标准线圈检出16例。

研究设计与样本

  • 回顾性研究,共纳入40例(18例女性,平均年龄67.0±15.1岁)头颈部肿瘤且临床怀疑骨侵犯的患者。
  • 所有患者均使用15通道牙科线圈与标准20通道线圈行MRI检查。
  • 两名放射科医生回顾性评估信噪比(SNR)和对比噪声比(CNR),并以5分评分量表主观评估骨侵犯、图像质量、金属伪影和肿瘤勾画。
  • 以组织病理学为参考标准,计算两种线圈检测骨侵犯的敏感度和特异度,采用Cohen's kappa评估观察者间一致性。

关键影像表现/征象

  • 15通道线圈肿瘤SNR为12.1,标准20通道线圈为9.9;15通道线圈肿瘤CNR为3.6,标准线圈为1.5,差异均有统计学意义。
  • 15通道线圈在图像质量、金属伪影、病灶勾画及骨侵犯评估方面均优于标准线圈。
  • 骨侵犯评估、整体图像质量和伪影减少的观察者间一致性极好,病灶勾画为显著一致。

诊断标准与关键数值

  • 15通道线圈检测骨侵犯敏感度100%,特异度97%;标准20通道线圈敏感度76%,特异度82%,均以组织病理学为参考标准。

临床应用与阅片要点

  • 对于临床怀疑骨侵犯的头颈部肿瘤患者,使用15通道牙科线圈可提高骨侵犯检出的敏感度和特异度,并改善图像质量、金属伪影和病灶勾画。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: To compare the diagnostic accuracy of a 15-channel dental coil with that of a standard 20-channel coil in the evaluation of head and neck tumours with suspected bone invasion.

METHODS AND MATERIALS: A total of 40 patients (18 females, mean age 67.0 ± 15.1 years) with head and neck cancer and clinical suspicion of bone invasion underwent MRI using both coils. Two radiologists retrospectively evaluated the images for signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR), subjectively for bone infiltration, image quality, metal artefacts and tumour delineation using a 5-point grading scale. The sensitivity and specificity of both coils in detecting bone invasion were calculated in histopathological correlation. Interobserver agreement was assessed using Cohen's kappa.

RESULTS: SNR and CNR values of tumours were significantly higher with the 15-channel coil (SNR: 12.1 vs. 9.9; CNR: 3.6 vs. 1.5). The 15-channel coil was superior in image quality, metal artifacts, lesion delineation, and assessment of bone infiltration. Interobserver agreement was very good for bone infiltration assessment, overall image quality and artefact reduction, and substantial for lesion delineation. The 15-channel coil was found to be more sensitive (100% vs. 76%) and more specific (97% vs. 82%) in detecting bone invasion. Twenty-three histopathologically confirmed tumours with bone invasion were correctly detected with the 15-channel coil, whereas sixteen were detected with the standard coil.

CONCLUSIONS: MRI with the 15-channel coil provides significantly better image quality and accuracy in the assessment of head and neck cancer, particularly in cases of bone invasion, compared with the standard coil.

CLINICAL RELEVANCE STATEMENT: The use of a 15-channel coil enhances diagnostic accuracy in the detection and characterization of head and neck cancer especially regarding bone infiltration and contributes to more individualized cancer treatment.

来源:Eur J Radiol,查看原文。版权归原作者所有。

引用

  1. Burck I, Liu X, Mahmoudi S, et al. MR imaging with 15-channel dental coil compared with standard 20-channel coil in patients with head and neck tumours and suspected bone invasion. Eur J Radiol. 2025. PMID: 41319371. doi:10.1016/j.ejrad.2025.112575

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。