三维超声中提肛肌裂孔的自动识别与测量:一项产后妇女的临床研究

Automated Recognition and Measurement for Levator Hiatus in 3D Ultrasound: A Clinical Study in Postpartum Women.
PubMed 文献 · J Ultrasound Med · 2025-11-25

本研究回顾性分析100例产后妇女的经会阴三维超声容积,评估基于卷积神经网络的AI软件自动识别并测量最小提肛肌裂孔的性能及其与手动测量的一致性。总体自动识别率为86.69%,不同功能状态下识别率不同;自动测量与手动测量在裂孔面积、前后径和左右径上一致性良好,但与提肛肌尿道间隙测量的一致性最差。作者认为该软件可行,可减少观察者间变异,但提肛肌尿道间隙测量仍需优化。

要点

  • 自动识别率在静息状态为94%,收缩状态为86.17%,Valsalva状态为79.80%。
  • 自动与手动测量在裂孔面积、前后径、左右径上ICC均大于0.75,一致性良好。
  • 提肛肌尿道间隙测量的自动与手动一致性最差。

核心结论

  • 基于卷积神经网络的自动软件在产后妇女中识别和测量提肛肌裂孔是可行的,可减少观察者间变异、标准化盆底评估并提高工作流程效率,但提肛肌尿道间隙测量需要优化以增强临床适用性。

研究设计与样本

  • 回顾性分析100例产后妇女的经会阴三维超声容积,使用SonoScape S60超声系统,由两名超声医师(一名初级、一名高级)独立手动测量,并与基于卷积神经网络的自动软件(Auto PF软件,SonoScape)测量结果进行比较。

诊断标准与关键数值

  • 手动测量包括前后径、左右径、提肛肌裂孔面积,在静息、最大盆底肌收缩动作和最大Valsalva动作时评估;提肛肌尿道间隙仅在收缩时测量。
  • 自动软件在自动识别失败时允许手动调整切面后重新尝试自动测量。
  • 总体自动识别率为86.69%,静息状态为94%,收缩状态为86.17%,Valsalva状态为79.80%。
  • 自动测量与手动测量在提肛肌裂孔面积、前后径和左右径上ICC均大于0.75,与高级超声医师测量在提肛肌裂孔面积和前后径上ICC均大于0.90。
  • 自动与手动测量的一致性在提肛肌尿道间隙测量中最差。

临床应用与阅片要点

  • 自动软件可用于产后妇女经会阴三维超声中提肛肌裂孔的识别与测量,有助于减少观察者间变异并标准化盆底评估。
  • 自动识别失败时,可手动调整切面后重新尝试自动测量。
  • 高级超声医师的手动测量与自动结果一致性更高。
  • 提肛肌尿道间隙的自动测量一致性最差,作者认为该测量需要优化以增强临床适用性。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: To evaluate the performance of a CNN-based (convolutional neural networks-based) AI software for automatic recognition and measurement of minimal levator hiatus in transperineal ultrasound volumes from postpartum women, and to assess its agreement with manual measurements in different functional states.

METHODS: We conducted a retrospective analysis of 100 transperineal ultrasound volumes measured independently by two sonographers (one junior, one senior) using a SonoScape S60 ultrasound system. Manual measurements included anteroposterior diameter (AP), left-to-right diameter (LR), levator hiatus area (HA), which were assessed at rest, during maximum pelvic floor contraction maneuver, and during maximum Valsalva maneuver. The levator-urethra gap (LUG) was additionally measured during contraction. The same volumes were identified and measured using automated CNN-based software (Auto PF software, SonoScape). When automatic identification failed, manual slice adjustment was permitted before reattempting automated measurement. The automatic recognition rate was recorded in different functional states. Inter-rater reliability between manual and automated measurements was evaluated using intraclass correlation coefficient (ICC) and Bland-Altman analysis.

RESULTS: The overall automatic recognition rate was 86.69%, varying by functional state: 94% at rest, 86.17% during contraction, 79.80% during Valsalva. Automated measurements showed good agreement with manual measurements for HA, AP and LR (ICC > 0.75), excellent agreement with the senior sonographer's measurement for HA, AP (ICC > 0.90). The senior sonographer's measurements demonstrated higher concordance with automated results. Furthermore, the agreement was poorest for LUG measurements.

CONCLUSIONS: Automated software is feasible for the recognition and measurement of LH in postpartum women, which can reduce the inter-observer variability, standardize pelvic floor assessment and improve workflow efficiency. Our study supports the clinical utility of CNN-based AI ultrasound software, but optimization is needed for LUG measurements to enhance clinical applicability.

来源:J Ultrasound Med,查看原文。版权归原作者所有。