比较多用途无针筒与传统一次性双筒高压注射器在增强CT中的效率、成本与技术员满意度

Comparing multi-use syringeless and conventional single-use dual-syringe injectors in contrast-enhanced CT: efficiency, cost, and technologist satisfaction.
PubMed 文献 · Eur Radiol · 2025-11-27

该回顾性研究比较多用途无针筒高压注射器(MU-SPI)与传统一次性双筒高压注射器(SU-DSPI)在增强CT中的操作效率、对比剂利用及技术员满意度。结果显示MU-SPI准备时间显著更短、对比剂浪费更少、成本更低且技术员评价更优,而图像质量无显著差异。

要点

  • MU-SPI平均准备时间51.0秒,显著短于SU-DSPI的198.8秒(p<0.001)。
  • SU-DSPI每例不可避免对比剂浪费24.6 mL,MU-SPI无此报告。
  • 两组对比增强质量无显著差异(p>0.5),MU-SPI成本更低($21.70 vs $25.00)。

核心结论

  • MU-SPI显著提高操作效率并降低成本,图像质量相当,同时提升技术员满意度。

研究设计与样本

  • IRB批准、HIPAA合规的回顾性研究,前瞻性记录随机选取的22例MU-SPI和30例SU-DSPI患者(平均年龄59 vs 60岁)的设备安装/释放时间;另用51对同一CT检查的配对病例评估增强质量;回顾性记录688例SU-DSPI(1例缺失)的对比剂浪费;研究期间(2021年7月至2022年3月)医疗系统共记录MU-SPI 9304例(平均年龄52岁±15;男4532例)和SU-DSPI 689例(平均年龄53岁±16;男354例)。

诊断标准与关键数值

  • 对比增强质量在MU-SPI与SU-DSPI之间无显著差异(p>0.5)。
  • MU-SPI每例平均准备时间51.0秒(SD 26.5),SU-DSPI为198.8秒(SD 26.4),p<0.001。
  • SU-DSPI每例不可避免对比剂浪费24.6 mL(SD 16.3)。
  • MU-SPI成本$21.70,SU-DSPI成本$25.00。
  • 技术员评价MU-SPI性能优于SU-DSPI(p=0.027)。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: To compare the operational efficiency, including performance, contrast media (CM) utilization, and technologist satisfaction between a multi-use syringeless power injector (MU-SPI) and a conventional single-use dual-syringe power injector (SU-DSPI).

MATERIALS AND METHODS: In this IRB-approved, HIPAA-compliant retrospective study, we prospectively recorded randomly selected cases from the first week of the caseload to document equipment setup/release times, including 22 patients (mean age 59) using MU-SPI and 30 patients (mean age 60) using SU-DSPI. In addition, the quality of contrast enhancement was evaluated using 51 randomly selected paired cases where subjects had the same CT study performed with both MU-SPI and SU-DSPI. Finally, we retrospectively recorded unavoidable CM wastage in 688 SU-DSPI cases (one missing). Overall, the healthcare system recorded 9304 cases (mean age, 52 years ± 15; 4532 men) using MU-SPI and 689 cases (mean age, 53 years ± 16; 354 men) using SU-DSPI in the study period from July 2021 to March 2022. Statistical analyses were performed using t-tests, Fisher's exact test, the Wilcoxon signed-rank test, and multivariate linear regression models.

RESULTS: Mean injector preparation time per examination was significantly lower for MU-SPI [51.0 (SD: 26.5) seconds] versus SU-DSPI [198.8 (SD: 26.4) seconds] (p < 0.001). SU-DSPI had unavoidable CM wastage of 24.6 (SD: 16.3) mL per exam. Contrast quality showed no significant differences between the groups (p > 0.5). Costs were lower with MU-SPI ($21.70 vs. $25.00). Finally, technologists rated the MU-SPI as having better performance compared to SU-DSPI (p = 0.027).

CONCLUSION: MU-SPI significantly enhances operational efficiency and reduces costs with comparable image quality, improving technologist satisfaction.

KEY POINTS: Question Evaluating how multi-use syringeless power injectors improve CT workflow efficiency and resource utilization amid increasing clinical demand for contrast-enhanced CT examinations. Findings The syringeless injector reduced preparation time, minimized contrast wastage, improved technologist satisfaction, and lowered consumable costs compared with the dual-syringe injector, without compromising image quality. Clinical relevance Implementing multi-use syringeless power injectors can streamline CT workflow and reduce costs. The saved time may enhance patient interaction and preparation, while improved efficiency and lower consumable expenses support overall CT department productivity.