优化膝关节MRI:三维质子密度涡轮自旋回波nDixon序列与二维高分辨率涡轮自旋回波在缩短一半采集时间下的比较
前瞻性研究纳入32例有症状受试者,在3T扫描仪上比较深度学习重建的二维TSE方案(4个序列,总时间12分6秒,作为参考标准)与三维nDixon TSE序列(5分55秒)。nDixon的骨、软骨、肌肉SNR更高,但关节液SNR和CNR低于TSE;两者对骨折、韧带撕裂、软骨或半月板病变的敏感度无差异(P≥0.23)。作者认为单个三维nDixon序列可能替代常规高分辨率TSE方案用于常规膝关节MRI。
要点
- 三维nDixon TSE序列采集时间5分55秒,二维TSE参考方案为12分6秒。
- nDixon的骨、软骨、肌肉SNR高于二维TSE,但关节液SNR和CNR低于二维TSE。
- 两者检测骨折、韧带撕裂、软骨或半月板病变的敏感度无显著差异(P≥0.23)。
核心结论
- 单个三维nDixon序列可能适合替代常规高分辨率TSE成像方案用于常规膝关节MRI。
研究设计与样本
- 前瞻性伦理批准研究,纳入32例有症状受试者,在临床3T扫描仪上行膝关节MRI;二维深度学习重建增强TSE方案(4个序列,总时间12分6秒)作为参考标准,三维深度学习重建增强nDixon TSE序列采集时间5分55秒。
关键影像表现/征象
- nDixon序列的骨、软骨和肌肉SNR高于常规TSE图像。
- TSE序列的关节液SNR高于nDixon图像(P<0.001)。
- TSE图像的CNR高于nDixon图像(P≤0.04)。
诊断标准与关键数值
- 检测骨折、韧带撕裂、软骨或半月板病变的敏感度在两种成像方案间无差异(P≥0.23)。
临床应用与阅片要点
- 单个三维nDixon序列可能替代常规高分辨率TSE方案用于常规膝关节MRI。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: The objective of this study was to evaluate the use of a single proton density (PD) new Dixon (nDixon) three-dimensional turbo spin echo (TSE) sequence for comprehensive knee joint assessment. This was achieved by comparing image quality and diagnostic performance with a conventional (PD + T1) two-dimensional TSE knee MRI protocol.
MATERIAL AND METHODS: This prospective ethics-approved study included 32 symptomatic subjects who underwent knee MRI on a clinical 3T scanner. Each subject was scanned using a two-dimensional deep-learning reconstruction (DLR)-enhanced TSE protocol, which involved four sequences and required a total time of 12 min and 6 s. This served as the reference standard. In addition, a DLR-enhanced three-dimensional nDixon TSE sequence was used, which required 5 min and 55 s. The signal-to-noise ratio (SNR) and contrast-to-noise ratios (CNR) of commonly depicted structures as well as common diagnostic findings in knee MRI were compared between the two imaging protocols.
RESULTS: The SNRs of bone, cartilage, and muscle were higher with the nDixon than with the conventional TSE images, whereas the SNR of joint fluid was higher with the TSE images (P < 0.001). This resulted in higher CNRs with the TSE than with the nDixon images (P ≤ 0.04). The sensitivities in detecting fractures, ligament tears, and cartilaginous or meniscal lesions did not differ between the two imaging protocols (P ≥ 0.23).
CONCLUSION: A single three-dimensional nDixon sequence may be a suitable replacement for a conventional high-resolution TSE imaging protocol in routine knee MRI.