文献速递腹部技术与安全MR

3.0T超短回波时间MR成像检测地中海贫血患者肝铁过载

Detection of Liver Iron Overload in Patients With Thalassemia With Ultra-short Echo Time MR Imaging on 3.0T.
PubMed 文献 · Invest Radiol · 2026-09-01

前瞻性纳入地中海贫血患者,对比1.5T梯度回波与3.0T超短回波时间序列定量肝铁浓度,58例可测患者中3.0T各序列与1.5T相关性高(R² 0.9701-0.9827),Bland-Altman偏倚小、一致性界限窄,15回波屏气方案表现最佳。

要点

  • 3.0T UTE各采集方案与1.5T GRE的R²为0.9701–0.9827,相关性高。
  • 3.0T UTE 15回波屏气方案在四种方案中表现最佳。
  • 可评估的肝铁浓度范围为1.3–39.5 mg/g,覆盖宽范围。

核心结论

  • 3.0T超短回波时间MRI序列可在地中海贫血患者中提供临床可行且准确的肝铁过载评估,肝铁浓度范围为1.3–39.5 mg/g,支持其用于该人群的肝铁过载检测和治疗决策。

研究设计与样本

  • 前瞻性纳入因肝铁评估行MRI的地中海贫血患者,每例均行1.5T GRE和3.0T UTE序列检查。
  • 3.0T UTE评估四种采集方案:7回波自由呼吸、7回波屏气、15回波自由呼吸、15回波屏气。
  • 共纳入63例,5例因1.5T MRI肝铁浓度过高无法测量被排除,最终分析58例。
  • 最终分析58例平均年龄34.3±16.1岁,男性24例(41.4%),重型地中海贫血42例(72.4%),规律输血31例(53.4%)。
  • 以1.5T GRE为参考标准,分析1.5T GRE与各3.0T UTE序列的相关性,并用Bland-Altman分析评估一致性。

诊断标准与关键数值

  • 3.0T UTE各序列与1.5T GRE的R²为0.9701–0.9827。
  • Bland-Altman分析显示偏倚极小、一致性界限窄。
  • 3.0T UTE 15回波屏气方案表现最强。

临床应用与阅片要点

  • 3.0T UTE 15回波屏气方案在四种方案中表现最佳,可作为肝铁定量采集方案的优先选择。
  • 3.0T UTE与1.5T GRE相关性高且一致性界限窄,支持在3.0T系统上用UTE进行肝铁过载检测和治疗决策。
  • 评估肝铁浓度范围覆盖1.3–39.5 mg/g,包括严重过载伴较低R2*值的情况。

局限性

  • 5例因1.5T MRI肝铁浓度过高无法测量而被排除。
  • 结论指出这些发现支持3.0T UTE MRI的临床效用,但原文未报告需进一步研究等自我限定。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: Accurate detection and quantification of liver iron overload (LIO) in patients with thalassemia is crucial for guiding iron chelation therapy and preventing iron-related organ damage. While conventional multiecho gradient-echo (GRE) based MR at 1.5T is the clinical standard, with increasing availability of 3.0T systems, clinically reliable alternatives are needed. The ultra-short echo time (UTE) MRI sequence may offer improved assessment of LIO on 3.0T. The objective of this study was to evaluate 3.0T UTE for assessing LIO in thalassemia patients and directly compared with the standard 1.5T GRE as reference, particularly at severe overload with lower R2* values.

MATERIALS AND METHODS: Patients with thalassemia referred for liver iron assessment by MRI were prospectively enrolled. Each participant underwent liver iron quantification using both 1.5T GRE and 3.0T UTE sequences. For the latter, 4 different acquisition protocols were assessed: 7-echo free breathing (3.0T UTE 7E FB), 7-echo breath-hold (3.0T UTE 7E BH), 15-echo free breathing (3.0T UTE 15E FB), and 15-echo breath-hold (3.0T UTE 15E BH). The correlation between 1.5T GRE and each UTE sequence was analyzed. The agreement was further assessed using Bland-Altman analysis.

RESULTS: Sixty-three patients were enrolled; 5 were excluded due to unmeasurably high liver iron concentration (LIC) by 1.5T MRI. The remaining 58 patients had a mean age of 34.3 ± 16.1 years; 24 (41.4%) were male, and 42 (72.4%) had thalassemia major. Regular transfusions were noted in 31 (53.4%). All 3.0T UTE sequences demonstrated excellent correlation with 1.5T GRE ( R2 , 0.9701-0.9827). Bland-Altman analysis indicated minimal bias and narrow limits of agreement. The 3.0T UTE 15E BH protocol yielded the strongest performance.

CONCLUSIONS: 3.0T UTE MRI sequences provide clinically feasible and accurate assessment of liver iron overload in thalassemia patients across a broad range of LIC values from 1.3 to 39.5 mg/g. These findings support the clinical utility of 3.0T UTE MRI for LIO detection and therapeutic decision-making in this population.

来源:Invest Radiol,查看原文。版权归原作者所有。

引用

  1. Krittayaphong R, Saiviroonporn P, Zhang S, et al. Detection of Liver Iron Overload in Patients With Thalassemia With Ultra-short Echo Time MR Imaging on 3.0T. Invest Radiol. 2026. PMID: 41261381. doi:10.1097/RLI.0000000000001261

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。