利用聚乙烯醇和三维打印模具开发模拟器官组织的超声成像体模的方法
研究用聚乙烯醇冷冻凝胶结合硅胶翻模与三维打印技术制作模拟兔肝和人甲状腺组织的超声体模,并评估其声学与力学特性。结果显示体模的声速、声阻抗、杨氏模量、剪切模量和密度与真实组织接近,并复现超声散斑模式。作者认为该体模成本低、组织复现准确,适用于超声成像研究、培训、教学和校准。
要点
- 聚乙烯醇冷冻凝胶体模可模拟兔肝与人甲状腺的声学与力学特性。
- 体模3 MHz声速分别为1528±6 m/s(兔肝)和1531±6 m/s(人甲状腺)。
- 体模复现超声散斑模式,增强成像应用的真实感。
核心结论
- 聚乙烯醇冷冻凝胶体模的声学与力学特性与兔肝和人甲状腺组织接近,其低成本制作和准确组织复现使其适用于超声成像研究、培训、教学和校准。
研究设计与样本
- 采用多次冻融循环制备聚乙烯醇冷冻凝胶体模,并用硅胶翻模与三维打印制作解剖学精确的模具以精确复现组织。
关键影像表现/征象
- 兔肝体模的声速为1528±6 m/s,人甲状腺体模的声速为1531±6 m/s(3 MHz)。
- 兔肝体模的声阻抗为1.60±0.01 Mrayl,人甲状腺体模的声阻抗为1.64±0.02 Mrayl。
- 兔肝体模的杨氏模量为129.02±3.43 kPa,人甲状腺体模的杨氏模量为131.48±5.78 kPa。
- 兔肝体模的剪切模量为43.01±3.43 kPa,人甲状腺体模的剪切模量为43.83±2.39 kPa。
- 兔肝体模的密度为1050 kg/m³,人甲状腺体模的密度为1070 kg/m³;体模复现了超声散斑模式。
诊断标准与关键数值
- 以3 MHz下声速、声阻抗、杨氏模量、剪切模量和密度为评估指标,体模数值与兔肝和人甲状腺组织接近。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: Medical imaging has advanced significantly, driving the need for realistic phantoms to enhance diagnosis, training and system calibration. However, existing homemade phantoms have limited lifespans and suboptimal tissue-mimicking properties, while commercial phantoms are often costly and inaccessible to many researchers. This study aims to fabricate polyvinyl alcohol cryogel phantoms replicating rabbit liver and human thyroid tissues using 3-D printing and silicone moulding techniques.
METHODS: Polyvinyl alcohol cryogel phantoms were fabricated using multiple freeze-thaw cycles to achieve the desired acoustic and mechanical properties. Silicone moulding and 3-D printing enable anatomically accurate moulds for precise tissue replication. The phantoms' acoustic and mechanical properties, including speed of sound, acoustic impedance, Young's modulus, shear modulus and density, were evaluated.
RESULTS: The fabricated phantoms exhibited acoustic (3 MHz) and mechanical properties closely related to those of actual rabbit liver and human thyroid tissues, including speed of sound (1528±6 m/s and 1531±6 m/s), acoustic impedance (1.60±0.01 Mrayl and 1.64±0.02 Mrayl), Young's modulus (129.02±3.43 kPa and 131.48±5.78 kPa), shear modulus (43.01±3.43 kPa and 43.83±2.39 kPa) and density (1050 kg/m³ and 1070 kg/m³), respectively. The phantoms also replicated ultrasound speckle patterns, enhancing their realism for imaging applications.
CONCLUSION: The polyvinyl alcohol cryogel phantoms closely resemble the acoustic and mechanical properties of rabbit liver and human thyroid tissues. Their cost-effective fabrication and accurate tissue replication make them ideal for ultrasound imaging research, training, education and calibration.