文献速递腹部技术与安全CT

腹部盆腔光子计数CT虚拟单能成像的窗设置界定:定量与定性评价

Defining window settings for abdominopelvic photon-counting CT virtual monoenergetic imaging: quantitative and qualitative evaluation.
PubMed 文献 · Eur J Radiol · 2025-11-17

回顾性分析39例患者双期腹部盆腔光子计数CT的60与70 keV虚拟单能图像,由三名放射科医师手动调整窗中心/窗宽以获得软组织与肝脏最佳显示。60 keV较70 keV平均对比噪声比提高24–27%(p<0.001);采用优化窗设置后,60 keV软组织与肝脏的显示对比噪声比及主观排名和印象评分均显著优于默认窗设置(p<0.001)。

要点

  • 60 keV虚拟单能图像较70 keV平均对比噪声比提高24–27%(p<0.001)。
  • 默认软组织窗下60 keV动脉期显示对比噪声比降低7%、静脉期升高19%;优化窗下分别升高51%和27%。
  • 默认肝脏窗下60 keV两期显示对比噪声比降低10–14%;优化窗下升高75–92%。

核心结论

  • 优化窗设置可显著改善光子计数CT虚拟单能图像上软组织和肝脏的显示,尤其在60 keV时。

研究设计与样本

  • 经IRB批准的回顾性分析,纳入39例患者,获取双期腹部盆腔光子计数CT的60和70 keV虚拟单能图像,由三名放射科医师手动调整窗中心/窗宽。

诊断标准与关键数值

  • 60 keV较70 keV平均对比噪声比提高24–27%(p<0.001)。
  • 默认软组织窗下,60 keV较70 keV显示对比噪声比在动脉期降低7%、静脉期升高19%;优化窗下动脉期升高51%、静脉期升高27%。
  • 默认肝脏窗下,60 keV较70 keV两期显示对比噪声比降低10–14%;优化窗下两期升高75–92%。
  • 60 keV图像采用优化窗设置时,软组织和肝脏评估的主观排名与印象评分均显著优于默认窗(p<0.001)。

临床应用与阅片要点

  • 对光子计数CT腹部盆腔60 keV虚拟单能图像,原文指出默认窗设置下图像较标准120 kV CT更亮,需手动调整窗设置。
  • 原文结论认为优化窗设置可显著改善60 keV虚拟单能图像上软组织和肝脏的显示。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

BACKGROUND: Virtual monoenergetic images (VMIs) at 60 keV are recommended for contrast-enhanced soft tissue imaging with decreased radiation dose on clinical photon-counting CT (PCCT). However, 60 keV VMIs appear brighter with default windowing compared to standard 120-kV CT, necessitating manual window adjustments.

OBJECTIVE: To determine optimal soft tissue and liver window settings for PCCT abdominopelvic imaging at 60 and 70 keV, and to quantitatively and qualitatively evaluate default versus optimal windowing.

METHODS: With IRB approval, retrospective analysis was performed on dual-phase abdominopelvic PCCT 60 and 70 keV VMIs from 39 patients. Three radiologists manually adjusted center/width (C/W) settings for optimal soft tissue and liver visualization. Contrast-to-noise ratio (CNR) was quantified for each phase and keV level, and display-CNR (dCNR) was quantified across multiple windows: PACS default and optimal windows for soft tissue and liver, and vendor-recommended soft tissue windowing. Readers ranked datasets viewed in different windows from most to least preferred and assigned an overall impression (optimal, acceptable, or poor). Expected ranking and impression scores were analyzed, with p-values < 0.05 indicating statistical significance.

RESULTS: Mean CNR increased by 24-27 % (p < 0.001) on 60 keV compared to 70 keV images. With default soft tissue windowing, dCNR at 60 keV decreased by 7 % (arterial) and increased by 19 % (venous) compared to 70 keV but increased by 51 % (arterial) and 27 % (venous) with optimal windowing. Liver dCNR decreased by 10-14 % at 60 keV using default windowing but increased by 75-92 % with optimal windowing on both phases. 60 keV images viewed at optimal windowing scored significantly better in terms of both ranking and impression compared to the default windowing for both soft tissue and liver evaluation on both phases (p < 0.001).

CONCLUSION: Optimized window settings significantly enhance soft tissue and liver visualization on PCCT VMIs, particularly at 60 keV.

来源:Eur J Radiol,查看原文。版权归原作者所有。