婴幼儿光子计数CT心脏成像中辐射剂量与图像质量的平衡:70 kV与120 kV协议的比较
该回顾性研究比较了体重低于10 kg婴幼儿在光子计数CT上行超高清心脏CT时,70 kV与120 kV两种协议的图像质量和辐射剂量。58例扫描中,两种协议主动脉对比噪声比和主观评分相似,但70 kV组辐射剂量显著更低(CTDI 0.13 mGy vs 0.25 mGy,P<0.001),不过70 kV协议无法获得高级光谱重建。
要点
- 体重<10 kg婴幼儿心脏光子计数CT中,70 kV协议辐射剂量显著低于120 kV(CTDI 0.13 vs 0.25 mGy,P<0.001)。
- 70 kV与120 kV协议主动脉对比噪声比和5分Likert主观评分无显著差异。
- 70 kV协议无法获得目前仅高kV协议才有的高级光谱重建后处理。
核心结论
- 70 kV协议在婴幼儿心脏CT中显示出降低辐射剂量的前景,图像质量与120 kV协议相似,但无法获得高级光谱重建。
研究设计与样本
- IRB批准的回顾性研究,纳入体重<10 kg患儿行光子计数CT超高清心脏CT扫描,比较70 kV(28例)与120 kV(30例)两种协议,共58例扫描,由两名读片者评估图像质量并比较辐射剂量。
关键影像表现/征象
- 主动脉对比噪声比70 kV组14.8 vs 120 kV组17.8,P=0.6。
- 主动脉主观Likert评分两组中位数均为5分,P=0.2。
诊断标准与关键数值
- 主观图像质量采用5分Likert量表评估,两组主动脉评分中位数均为5分。
- 辐射剂量以CTDI比较:70 kV组0.13 mGy vs 120 kV组0.25 mGy,P<0.001。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND: Photon-counting CT (PCCT) has the potential to improve diagnosis and reduce radiation exposure in pediatric congenital heart disease. Early data in pediatric patients has shown advantages of advanced post-processing techniques that are currently only available with high kilovolt (kV) protocols.
OBJECTIVE: The purpose of our study was to compare image quality and radiation dose for low kV (70) and high kV (120) protocols in patients weighing under 10 kg.
METHODS: In this IRB-approved retrospective review of ultra-high pitch cardiac CT scans in patients under 10 kg, we evaluated two protocols on a PCCT scanner: low kV (70) and high kV (120). Patient demographics, contrast-to-noise ratio in aorta and pulmonary artery, and subjective image quality scores using a 5-point Likert scale were evaluated by two readers in addition to radiation dose comparison.
RESULTS: Fifty-eight scans were included, 28 in the 70 kV cohort and 30 in the 120 kV cohort. Median age was 0.17 years for 70 kV and 0.33 years for 120 kV and weight was 4.5 kg for 70 kV and 5.4 kg for 120 kV. Contrast-to-noise ratios (aorta 14.8 vs 17.8, P=0.6) and Likert scores for aorta (median 5 for both groups, P=0.2) were similar for both protocols. Radiation dose was significantly lower in the 70 kV group with CTDI of 0.13 mGy vs 0.25 mGy for 120 kV (P<0.001).
CONCLUSION: A 70 kV protocol shows promise for reducing radiation dose in infants with similar image quality compared to the 120 kV protocol, although advanced spectral reconstructions would not be available.