光子计数CT虚拟单能成像检测心肌晚期碘增强与心脏MRI的对比研究
该研究基于前瞻性研究队列的事后分析,纳入27例同日行光子计数探测器CT延迟增强与晚期钆增强MRI的参与者,评估40-90 keV不同虚拟单能级图像质量与心肌纤维化诊断效能,以MRI为参考。50 keV在客观与主观图像质量间取得平衡,在逐节段分析中检测局灶性纤维化的敏感度87.4%、特异度97.8%、准确率95.9%,与MRI一致性κ=0.86。
要点
- 50 keV虚拟单能级在图像噪声、对比噪声比与主观质量间取得平衡,诊断效能最高。
- 50 keV逐节段检测局灶性纤维化敏感度87.4%、特异度97.8%、准确率95.9%,与MRI一致性κ=0.86。
- 50 keV重建在区分纤维化模式方面与MRI一致性最高。
核心结论
- 光子计数探测器CT心肌延迟增强采用50 keV虚拟单能重建是检测和表征心肌纤维化的有价值工具,与其他keV水平相比诊断效能最佳且与MRI一致性最高。
研究设计与样本
- 前瞻性研究队列的事后亚组分析,纳入2021年7月至2022年1月间接受晚期钆增强MRI并同日行光子计数探测器CT延迟增强的27例参与者(平均年龄52.9岁±17.2[SD],14例女性);CT延迟增强扫描在40-90 keV不同虚拟单能级下由两名盲法阅片者评估图像质量及心肌纤维化存在与分布,计算诊断效能并以Cohen κ评估与MRI一致性。
诊断标准与关键数值
- 以MRI为参考,27例中12例(44%)参与者及459个节段中87个(19%)节段晚期钆增强MRI阳性;50 keV虚拟单能级在逐节段分析中检测局灶性纤维化的敏感度为87.4%(95%CI:80.4,94.3)、特异度为97.8%(95%CI:96.4,99.3)、准确率为95.9%(95%CI:94.0,97.7),与MRI一致性κ=0.86。
临床应用与阅片要点
- 作者认为光子计数探测器CT心肌延迟增强50 keV虚拟单能重建是检测和表征心肌纤维化的有价值工具,与其他keV水平相比诊断效能最佳且与MRI一致性最高。
局限性
- 原文未报告局限性
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
Purpose To investigate the image quality and diagnostic performance of photon-counting detector CT-based delayed enhancement (DE) at different virtual monoenergetic imaging (VMI) levels in detecting and characterizing myocardial fibrosis, using MRI as the reference. Materials and Methods This post hoc subanalysis of a prospective research cohort included participants who underwent late gadolinium enhancement MRI between July 2021 and January 2022 and same-day research CT DE on a photon-counting detector CT system. CT DE scans at different VMI levels (40-90 keV) were evaluated by two blinded readers in terms of subjective and objective image quality and presence and distribution of myocardial fibrosis. Diagnostic performance measures were calculated, and agreement with MRI was evaluated with Cohen κ statistics. Results The CT DE scans of 27 participants (mean age, 52.9 years ± 17.2 [SD]; 14 female) were analyzed. Twelve of 27 (44%) participants and 87 of 459 (19%) segments had a positive late gadolinium enhancement MRI. VMI at 50 keV demonstrated an adequate trade-off between objective (image noise and contrast-to-noise ratio) and subjective (image noise and sharpness and overall quality) image quality, exhibiting the highest sensitivity, specificity, and accuracy in focal fibrosis detection, in per-segment analyses (87.4% [95% CI: 80.4, 94.3], 97.8% [95% CI: 96.4, 99.3], and 95.9% [95% CI: 94.0, 97.7], respectively), with a strong agreement with MRI (κ = 0.86). These reconstructions showed the highest concordance in discriminating fibrosis patterns. Conclusion Photon-counting detector myocardial CT DE with VMI reconstructions at 50 keV represents a valuable tool for myocardial fibrosis detection and characterization, with best diagnostic performance and agreement with MRI when compared with other kiloelectron volt levels. Keywords: Virtual Monoenergetic Images, Late Iodine Enhancement, Myocardial Fibrosis, Photon-counting Detector CT Supplemental material is available for this article. © RSNA, 2025.