验证一种用于PET/CT引导介入的实用手动PET肿瘤大小描绘方法的准确性
回顾性纳入63例患者105个FDG高摄取肝/肺肿瘤,验证无需专用软件的手动窗宽法在PET上描绘肿瘤大小的准确性,并与MRI/CT解剖影像比较。PET与解剖影像的组内相关系数分别为0.92和0.93,总体平均大小差异为-1.6 mm(P<.001),作者认为该方法可提供与MRI和CT相当的肝肺肿瘤大小描绘且读者间相关性高。
要点
- 手动窗宽法在PET上描绘肝肺肿瘤大小与MRI/CT解剖影像的读者间组内相关系数高(PET 0.92,解剖影像0.93)。
- PET与解剖影像的总体平均肿瘤大小差异为-1.6 mm(P<.001),肝脏肿瘤为-2.0 mm(P<.001),肺肿瘤为-0.6 mm(P=.41)。
- 肿瘤≤2 cm的平均大小差异为-0.2 mm,>2 cm为-2.6 mm。
核心结论
- 无需专用软件的手动窗宽法在PET肿瘤图像上可提供与MRI和CT相当的肝肺肿瘤大小描绘,且读者间相关性高。
研究设计与样本
- 回顾性研究,纳入2017至2021年间63例患者共105个肿瘤(75个肝脏、30个肺部),要求FDG高摄取、在PET/CT与MRI(肝肿瘤)或CT(肺肿瘤)上均可见、肿瘤直径0.8–6.0 cm、且解剖影像在PET/CT后7天内完成;采用线性混合效应模型分析读者间一致性和PET与解剖测量的配对差异。
诊断标准与关键数值
- 读者间组内相关系数:PET测量为0.92,解剖影像测量为0.93;PET与解剖影像的总体平均肿瘤大小差异为-1.6 mm(P<.001),肺肿瘤为-0.6 mm(P=.41),肝肿瘤为-2.0 mm(P<.001),肿瘤≤2 cm为-0.2 mm,>2 cm为-2.6 mm。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
AIM: Positron emission tomography/computed tomography (PET/CT) guidance for tumour ablation procedures offers unique advantages, including intraprocedural margin assessment and visualisation of tumours not visible with other modalities. However, accurate tumour size depiction on PET is necessary. This paper validates a manual method for accurately depicting liver and lung tumour sizes on F18-fluorodeoxyglucose (FDG) PET compared to anatomic imaging with magnetic resonance imaging (MRI) and CT.
MATERIALS AND METHODS: This retrospective study included 63 patients with 105 tumours (75 liver and 30 lung) imaged between 2017 and 2021. Inclusion required FDG-avid tumours in the liver or lung, visibility on both PET/CT and MRI (liver tumours) or CT (lung tumours), tumour diameter 0.8-6.0 cm, and anatomical imaging within 7 days of PET/CT. Tumours were windowed on PET images using a manual segmentation method requiring no special software. Interreader agreement and pairwise difference between PET and anatomical tumour measurements were studied using linear mixed-effects models.
RESULTS: Intraclass correlation coefficients for interreader agreement were 0.92 for PET and 0.93 for anatomical tumour measurements. The overall average tumour size difference between PET and anatomical imaging was -1.6 mm (P<.001). The average size difference was -0.6 mm for lung tumours (P=.41) and -2.0 mm for liver tumours (P<.001). The average size difference for tumours ≤2 cm was -0.2 mm and -2.6 mm for tumours >2 cm.
CONCLUSION: A manual windowing method for PET tumour images provided comparable liver and lung tumour size depictions compared to MRI and CT with high interreader correlation.
引用
- Hu EY, Guo Y, Alencar RO, et al. Validation of a practical manual method of accurate positron emission tomography (PET) tumour size depiction for positron emission tomography/computed tomography (PET/CT)-guided interventions. Clin Radiol. 2025. PMID: 41297155. doi:10.1016/j.crad.2025.107153
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。